4.89M

82-01.54.455736-1.2 i15 Service Manual-EN

1.

EDAN Agile PLM Electronic Signature Information
--Signatures related to this document and performed in EDAN Agile PLM.
文件名称 i15血气生化分析仪维修手册 英文
文件编号 01.54.455736
版本 1.2
产品型号 BG5
项目编码(Project Code) 3300L-6000
签批信息:
作者 : 龚 静 (gongjing)
2018-09-07 14:32:00
审核人 : 谢 瑛瑛 (xieyingying)
审核人 : 王 芳 (wangfang)
2018-09-07 18:00:04
2018-09-10 09:09:04
审核人 : 陈 真真 (chenzhenzhen)
审核人 : 李 国霞 (liguoxia)
2018-09-10 08:36:46
2018-09-12 16:59:43
审核人 : 黄 高祥 (huanggaoxiang)
审核人 : 陈 晓东 (chenxiaodong)
2018-09-10 10:20:42
2018-09-07 16:53:08
审核人 : 章 星 (zhangxing)
2018-09-10 10:36:03
审核人 : 刘 桂林 (liuguilin)
2018-09-10 09:08:02
批准人 : 戴 颖 (daiying2)
2018-09-20 15:29:05
批准人 : 乐 秋红 (yueqiuhong)
批准人 : 郑 丽丽 (zhenglili)
2018-09-19 14:06:44
2018-09-19 22:58:10
批准人 : 潘 昭杨 (panzhaoyang)
2018-09-19 10:51:28
版权©深圳市理邦精密仪器股份有限公司 (Copyright©Edan Instrument,Inc.)

2.

3.

About this Manual
P/N:01.54.455736
MPN: 01.54.455736012
Release Date: August, 2018
© Copyright EDAN INSTRUMENTS, INC. 2016-2018. All rights reserved
Statement
This manual will help you to better understand the operation and maintenance of the product. It
is reminded that the product should be used strictly in compliance with this manual. User
operation failing to comply with this manual may result in malfunction or accident for which
EDAN INSTRUMENTS, INC. (hereinafter called EDAN) can not be held liable.
EDAN owns the copyrights of this manual. Without prior written consent of EDAN, any
materials contained in this manual shall not be photocopied, reproduced or translated into other
languages.
Materials protected by the copyright law, including but not limited to confidential information
such as technical information and patent information contained in this manual, shall not be
disclosed to any irrelevant third party.
The user shall understand that nothing in this manual grants him, expressly or implicitly, any
right or license to use any of the intellectual properties of EDAN.
EDAN holds the rights to modify, update, and ultimately explain this manual.
Responsibility of the Manufacturer
EDAN only considers itself responsible for any effect on safety, reliability and performance of
the equipment if:
Assembly operations, extensions, re-adjustments, modifications or repairs are carried out by
persons authorized by EDAN, and
The electrical installation of the relevant environment complies with national standards, and
The instrument is used in accordance with the instructions for use.
EDAN will make available on request circuit diagrams, component part lists, descriptions,
calibration instructions, or other information that will assist service personnel to repair those
I

4.

parts of the equipment that are designated by EDAN as repairable by service personnel.
Terms Used in this Manual
This guide is designed to give key concepts on safety precautions.
WARNING
A WARNING label advises against certain actions or situations that could result in personal
injury or death.
CAUTION
A CAUTION label advises against actions or situations that could damage equipment, produce
inaccurate data, or invalidate a procedure.
NOTE
A NOTE provides useful information regarding a function or a procedure.
II

5.

Table of Contents
1 Warranty and Service ................................................................................................................. 1
2 Safety Guide................................................................................................................................. 4
2.1 Introduction ......................................................................................................................... 4
2.2 Warnings and Cautions ....................................................................................................... 4
2.3 General Information ............................................................................................................ 6
3 Installation ................................................................................................................................... 7
3.1 Unpacking Inspection ......................................................................................................... 7
3.2 Installation Requirements ................................................................................................... 7
3.2.1 Environmental Requirements .................................................................................. 7
3.2.2 Power Requirements ................................................................................................ 8
3.3 Installing the Analyzer ........................................................................................................ 8
3.4 Connecting to AC Power .................................................................................................... 8
3.5 Replacing a Calibrant Fluid Pack........................................................................................ 8
4 Functional Checks ..................................................................................................................... 12
4.1 Switching On Check ......................................................................................................... 12
4.2 LCD Screen Check............................................................................................................ 12
4.3 Touch Screen Check ......................................................................................................... 12
4.4 Diagnostics ........................................................................................................................ 12
4.5 External Simulator Test .................................................................................................... 14
4.6 Control Test....................................................................................................................... 17
5 Maintenance .............................................................................................................................. 24
5.1 Maintenance Inspection .................................................................................................... 24
5.2 Maintenance of the Analyzer ............................................................................................ 24
6 Principle Introduction .............................................................................................................. 25
6.1 Measurement Method ....................................................................................................... 25
6.2 Determination of Test Results........................................................................................... 26
6.2.1 Determination of the Analyte Concentration ......................................................... 26
6.2.2 Determination of Cell Concentration ..................................................................... 27
III

6.

6.3 Equations for Calculated Parameters ................................................................................ 27
6.4 System Principle Block Diagram ...................................................................................... 31
7 Troubleshooting ........................................................................................................................ 32
7.1 Booting Failures ................................................................................................................ 32
7.2 Display Failures ................................................................................................................ 33
7.3 Touch Screen Failures ....................................................................................................... 33
7.4 Operation Failures ............................................................................................................. 34
7.5 Printer Failures .................................................................................................................. 35
7.6 Network Failures ............................................................................................................... 35
7.7 Diagnosis Database ........................................................................................................... 36
7.7.1 Searching for Diagnosis Data ................................................................................ 36
7.7.2 Viewing Details of Diagnosis Data ....................................................................... 37
7.7.3 Exporting/Printing Diagnosis Data ........................................................................ 38
8 Module’s Malfunction Verification ......................................................................................... 39
8.1 Verification of Main Control Board’ Circuit Failure ........................................................ 39
8.2 Verification of the Power Board’ Circuit Failure ............................................................. 40
8.4 Verifying Malfunction of the Touch Screen Drive Board ................................................ 41
8.5 Verifying Malfunction of the Scanner .............................................................................. 42
9 Disassembling the Analyzer ..................................................................................................... 43
9.1 Tools Required .................................................................................................................. 43
9.2 Disassembling the Main Unit ............................................................................................ 43
9.3 Disassembling the Lower Assembly ................................................................................. 44
9.3.1 Replacing the Main Control Board ........................................................................ 44
9.3.2 Replacing the Wi-Fi Board .................................................................................... 45
9.3.3 Replacing the Chamber Door Detection Board ..................................................... 46
9.3.4 Replacing the Ambient Temperature Detection Board.......................................... 47
9.3.5 Replacing the Analog Board .................................................................................. 47
9.3.6 Replacing the Cam Machine .................................................................................. 51
IV

7.

9.3.7 Replacing the Power Board ................................................................................... 56
9.3.8 Replacing the Power Interface Board .................................................................... 57
9.3.9 Replacing the Exhaust Fan .................................................................................... 58
9.3.10 Replacing the Speaker ......................................................................................... 59
9.3.11 Replacing the Bar Code Scanner ......................................................................... 59
9.3.12 Replacing the Spring Located in Test Cartridge Base ......................................... 61
9.3.13 Cleaning and Replacing Vacuum Pump .............................................................. 63
9.3.14 Replacing FPC Cable ........................................................................................... 64
9.3.15 Replacing the Cable of Temperature Detection Board ........................................ 72
9.3.16 Replacing and Cleaning Needle ........................................................................... 75
9.4 Disassembling the Upper Assembly ................................................................................. 77
9.4.1 Replacing the Touch Screen Drive Board ............................................................. 77
9.4.2 Replacing the LCD ................................................................................................ 78
9.4.3 Replacing the Touch Screen .................................................................................. 79
9.4.4 Replacing the Print Controller Board .................................................................... 80
9.4.5 Replacing the Printer Head .................................................................................... 81
10 Common Failures and Solutions ............................................................................................ 82
11 Replaceable Parts .................................................................................................................... 97
V

8.

i15 Blood Gas and Chemistry Analysis System Service Manual
Warranty and Service
1 Warranty and Service
Standard Service
EDAN provides a one-year-warranty for the warranted products. The warranty period begins on
the date the products are shipped to customers. If a customer promptly notifies EDAN of
customer’s warranty claim hereunder, EDAN will repair, adjust or replace (with new or exchange
replacement parts) EDAN’s products. EDAN warrants that any service it provides to customers
will be performed by trained individuals in a workmanlike manner.
Limitation of Warranty
Direct, indirect or final damage and delay caused by the following situations for which EDAN is
not responsible may void the warranty:
Groupware is dismounted, stretched or redebugged.
Unauthorized modification or misuse.
Damage caused by operating beyond the environmental specifications for the medical
product.
Change or remove original serial number label or Manufacturer symbol.
Improper use.
Service Procedure
(1) Fill in the Service Claim Form (SCF).
Fill in the SCF with detailed information including: Model Name, Serial Number (SN) and
Problem Phenomena.
EDAN should not have any obligation to take over the case without this information. The
form can be downloaded at: http://www.edan.com.cn or obtained from EDAN’s Service
Department.
(2) Send EDAN the SCF and Select a Solution.
Once the service department receives the fully filled SCF, EDAN’s engineer will offer a
solution in three working days. EDAN will follow out the case based on the two conditions
below:
-1-

9.

i15 Blood Gas and Chemistry Analysis System Service Manual
Warranty and Service
Within Warranty:
There are two options:
i) After receiving the Return Material Authorization (RMA) form from EDAN service
department, the customer sends EDAN the defective parts and informs about the shipment
tracking number. Then we will dispatch new part(s) to your confirmed address with
confirmed shipping invoice.
ii) The customer signs the Declaration Form and sends it back by email or fax. This form is
legally certificated to make sure the customer or end-user will return the defective parts to
EDAN on time. We will, at this option, dispatch the replacement one(s) with confirmed
shipping invoice.
NOTE:
Both Return Material Authorization Form and Declaration Form are offered by
EDAN service department once the SCF is confirmed by a service engineer.
The customer is responsible for freight & insurance charges when the equipment
is shipped to EDAN for service, including custom charges. EDAN is responsible
for the freight, insurance & custom charges from EDAN to the customer.
Out of Warranty:
After receiving the RMA form from the service department, the customer sends defective
parts to EDAN in advance. We will analyze the problems and discuss with the customer
about either repairing or replacing the part(s). Once the maintenance fee is invoiced and paid,
we will dispatch good part(s) to the confirmed address.
NOTE:
The customer is responsible for any freight & insurance charge for the returned
product.
(3) Obtain the RMA Form
Before the shipment of the materials, the customer must obtain an RMA form from our
service department, in which the RMA number, description of returning parts and shipping
instructions are included. The RMA number should be indicated on the outside of the
shipping container.
NOTE:
EDAN should not have any obligation to the end-user or customer who returns the
-2-

10.

i15 Blood Gas and Chemistry Analysis System Service Manual
Warranty and Service
goods without the notification by EDAN’s service department. The sender takes full
responsibility for the accounted fee.
(4) Send the Parts to EDAN.
Follow these recommended instructions:
Please disassemble the parts with anti-static facility, and never touch the parts with naked
hand.
Please pack the parts safely before return.
Please put the RMA number on the parcel.
Please describe the returned parts as ‘sample of *****’ and put the total value on the invoice,
and note on the invoice as ‘sample, no commercial value’.
Please confirm the invoice with EDAN before shipment.
Please send back the parts after EDAN’s confirmation.
Contact Information
If you have any question about maintenance, technical specifications or malfunctions of devices,
do not hesitate to contact us.
EDAN Instruments, Inc.
TEL: +86-755-26898321, 26899221
FAX: +86-755-26882223, 26898330
E-mail: support@edan.com.cn
-3-

11.

i15 Blood Gas and Chemistry Analysis System Service Manual
Safety Guide
2 Safety Guide
2.1 Introduction
This service manual is a reference for periodic preventive maintenance and corrective service
procedures for the i15 blood gas and chemistry analysis system. It provides information on
troubleshooting, assembly procedures, and instructions for functional testing as well as
performance verification. The manual is intended for use only by technically qualified service
personnel.
WARNING
Strictly follow the instructions as described in this manual during service. Otherwise the
system may be damaged, or personal injury may occur.
2.2 Warnings and Cautions
To avoid possible injury, please read the following warnings and cautions during the operation of
the system.
WARNING
1. The system is not intended for treatment.
2. The system is not intended for home use.
3. The system is intended for use only by trained technologists, nurses, physicians and
therapists.
4. SHOCK HAZARD – Do not attempt to connect or disconnect a power cord with wet
hands. Make sure your hands are clean and dry before touching a power cord.
5.
The sample is contained in the test cartridge, so test cartridges should be
disposed of as biohazardous waste, complying with local regulatory guidelines.
6. The system should be serviced only by authorized and qualified personnel. EDAN
does not assume any responsibility for damage or injury if modifications or repairs
are carried out by unauthorized personnel.
7. Service personnel should be familiar with the operation of the system. Refer to i15
Blood Gas and Chemistry Analysis System User Manual for details.
-4-

12.

i15 Blood Gas and Chemistry Analysis System Service Manual
Safety Guide
WARNING
8. Use and replace the substitutive parts provided or recommended by EDAN only.
9. Do not remove the battery while AC power is on.
10. Do not directly connect the battery to an electrical outlet.
11. Do not directly solder the lead wire or the battery terminal.
12. Do not touch the stitch of a calibrant fluid pack or the stitch of a vacuum pump.
13. Do not touch the roller or gear of a printer.
14. Do not touch the vane of an exhaust fan.
15.
Never use an external electronic simulator under electromagnetic
environment, and never touch it by hand during an external simulator test.
CAUTION
1. The system is designed for continuous operation. Avoid splashing water over the
system.
2. Use this system at a temperature between 10°C and 31°C. Outside this range, the
system may produce inaccurate results.
3. Do not use the system in a dusty environment with bad ventilation or in the presence
of corrosives.
4. Do not operate the system rudely, or the system may produce incorrect test results.
5. Make sure that there is no intense electromagnetic interference source around the
analyzer, such as radio transmitters, mobile phones, etc. Attention: large medical
electrical equipment such as electrosurgical equipment, radiological equipment and
magnetic resonance imaging equipment is likely to bring electromagnetic
interference.
6. Do not use the analyzer after its life cycle, and it should be disposed of according to
local regulations after its life cycle.
-5-

13.

i15 Blood Gas and Chemistry Analysis System Service Manual
Safety Guide
2.3 General Information
Comply with
Degree of protection against
harmful ingress of water
Degree of safety in presence of
flammable gases
EMC
Mode of operation
EN 61010-1: 2001, EN 61010-2-101: 2002,
EN 61326-1: 2006, EN 61326-2-6: 2006
Ordinary equipment (Sealed equipment without waterproof)
Equipment not suitable for use in presence of flammable gases
CISPR 11 Group 1, Class A
Continuous
-6-

14.

i15 Blood Gas and Chemistry Analysis System Service Manual
Installation
3 Installation
WARNING
To avoid electrical shock, do not remove any cover or board. Should problems or
malfunctions occur, unplug the power cord and power adaptor. Only qualified service
personnel should carry out servicing and troubleshooting.
3.1 Unpacking Inspection
Visually examine the package prior to unpacking. If there are any signs of mishandling or
damage, contact the carrier to claim for damage. After unpacking the device, customers should
follow the PACKING LIST to check the product carefully and to make sure no damage occurred
during transportation. Then, install the device according to the installation requirements and
procedures. If there is any problem, contact the manufacturer or the authorized local distributors
immediately.
3.2 Installation Requirements
3.2.1 Environmental Requirements
Location is of great importance for the smooth running of your analyzer. Prior to installing the
analyzer, choose a site that meets the following requirements:
Convenient for the analyzer to be connected to a grounded electrical receptacle in case it is
powered by AC power.
Keep the analyzer away from direct exposure to strong sunlight.
Ambient Temperature between 10°C and 31°C.
Relative Humidity of 25% - 80% (non-condensing).
Ambient Pressure within 70 - 106.6kPa (525 - 800mmHg).
Placed onto a clean and flat surface with good ventilation.
Keep the analyzer away from equipment with strong electric field and strong magnetic field.
Keep the analyzer away from explosive gases or vapors.
NOTE:
The requirements above also apply when your analyzer is powered by a rechargeable
lithium battery.
-7-

15.

i15 Blood Gas and Chemistry Analysis System Service Manual
Installation
3.2.2 Power Requirements
The analyzer needs to be connected to a grounded electrical outlet with the voltage between
100±10% VAC - 240±10% VAC and the frequency of 50/60Hz.
3.3 Installing the Analyzer
Please place the analyzer on a secure table surface with environments that meet the requirements
as described in section 3.2.
3.4 Connecting to AC Power
1.
Insert the power adaptor into the power connector on the analyzer.
2.
Plug the power cord into the power adaptor.
3.
Plug the power cord into a grounded electrical outlet.
NOTE:
Make sure the power requirements as described in 3.2.2 are met.
To avoid the analyzer and other electronic devices being damaged by electrical
power spikes, a surge protector is recommended.
3.5 Replacing a Calibrant Fluid Pack
WARNING
1. Never replace a calibrant fluid pack when the analyzer is off.
2. A calibrant fluid pack is intended for single use only. If a calibrant fluid pack is
removed from the system, it can not be inserted into the system again.
Follow the steps below to replace a calibrant fluid pack:
1.
Examine the expiration date on the package of a calibrant fluid pack to ensure it has not
expired.
2.
Remove the calibrant fluid pack from its package, and equilibrate it to room temperature.
The calibrant fluid pack needs to stand at room temperature for at least 24 hours.
3.
Wipe any moisture from the foil pouch with a dry clean cloth.
-8-

16.

i15 Blood Gas and Chemistry Analysis System Service Manual
4.
On the Main screen, press
5.
Press
Installation
to go to the System Setup screen.
. The system will go to the screen below:
Figure 3-1 Open Chamber Door
6.
Open the foil pouch, and remove the calibrant fluid pack from it.
NOTE:
Avoid tearing the bar code on the foil pouch.
7.
Remove the calibrant fluid pack cap, and remove the key by pushing it with your finger in
the direction as shown by the arrow in the picture below.
Key
Valve
NOTE:
-9-

17.

i15 Blood Gas and Chemistry Analysis System Service Manual
Installation
Avoid pushing the valve when removing the key.
8.
Unlock the calibrant fluid pack chamber lock with the key, and then pull the calibrant fluid
pack ejector to open the calibrant fluid pack chamber door.
9.
Remove the used calibrant fluid pack from the system. The system will go to the screen for
the next procedure, and the scanner will be turned on automatically.
Figure 3-2 Remove Old Calibrant Fluid Pack
10. Scan the bar code on the new calibrant fluid pack foil pouch with the bar code scanner.
If the bar code is scanned successfully, the system will beep and the scanner will be turned
off automatically. If the scanned data is valid, the system will display the screen for the next
procedure. If the scanned data is invalid, a message will pop up to prompt you.
If the scanner is turned off automatically, press
first, and then scan the bar code.
Figure 3-3 Scan Bar Code
11. Insert the new calibrant fluid pack into its chamber, and push it gently to make sure that it
- 10 -

18.

i15 Blood Gas and Chemistry Analysis System Service Manual
clicks into place. The system will go to the screen for the next procedure.
Figure 3-4 Install Calibrant Fluid Pack
12. Close the chamber door.
Figure 3-5 Close Chamber Door
13. Lock the calibrant fluid pack chamber lock to close the chamber door securely.
14. Press OK in the pop up message. The system will then go to the Main screen.
Figure 3-6 Replacement Success
- 11 -
Installation

19.

i15 Blood Gas and Chemistry Analysis System Service Manual
Functional Checks
4 Functional Checks
This section describes the procedure of a complete functional test to support recommended
preventive-maintenance schedules.
It is unnecessary to open the analyzer case for functional checks.
WARNING
Functional checks should be performed by service personnel whenever installation,
servicing, or troubleshooting is performed.
4.1 Switching On Check
Press the On/Off button on the left side of the analyzer to turn it on.
Check if the power indicator lights up, and if the screen lights up in a few seconds.
If any failure occurs, refer to section 7.1 Booting Failures for details.
4.2 LCD Screen Check
Observe if some characters are missing, or if there are bright spots or dark shadows on the LCD
screen. Observe if fonts and symbols displayed on the LCD screen are normal.
If any failure occurs, refer to section 7.2 Display Failures for details.
4.3 Touch Screen Check
Touch any available button on the touch screen to check if it is working properly.
If you touch the center of a button but the analyzer does not execute the right operation, calibrate
the touch screen. Refer to section 4.4 Dignostics for details.
If any operation failure occurs, refer to section 7.3 Touch Screen Failures for details.
4.4 Diagnostics
This function let you view the information for some modules, check the functioning of the pump,
valve, printer, and internal electronic simulator, calibrate the touch screen, export data, reset the
system password, reset the system setup to the factory setup, online update the analog board,
- 12 -

20.

i15 Blood Gas and Chemistry Analysis System Service Manual
Functional Checks
scan the system ID bar code, view software running logs, and clear data stored in the analyzer. It
helps to troubleshoot.
NOTE:
Only service engineers and engineers from the manufacturer can perform this action.
Figure 4-1 Diagnostics Screen
Follow the steps below to perform diagnostics:
1.
Press
to go to the Diagnostics screen.
2.
Refer to the table below to perform actions.
No.
To
Press
View the analog board information. Analog board software
version (it should be the latest version), front end board ID (it
should be consistent with the tracing list), ambient
1.
temperature (it should be 10 C - 35 C), ambient pressure (it
should be 730mmHg - 760mmHg), the heater 1 temperature
Analog Board
(it should be 10 C - 35 C), and the heater 2 temperature (it
should be 10 C - 35 C).
View the power board information. Power board software
version (it should be the latest version), 5V voltage (it should
be 4.5V - 5.5V), 12V voltage (it should be 11.5V - 12.5V),
2.
printer voltage (it should be 12V - 20V), input current (it
should be smaller than 1A), power board temperature (it
should be 10 C - 50 C), the rate of the exhaust fan (it should
be 0 – 8000 cycles/min, and for temperature below 29 C, the
exhaust fan does not run).
- 13 -
Power Board

21.

i15 Blood Gas and Chemistry Analysis System Service Manual
3.
Functional Checks
View the main control board information. Main control board
Main Board
software version (it should be the latest version).
4.
Check whether the pump is working properly.
Pump
5.
Check whether the valve is working properly.
Valve
Check whether the printer is working properly.
The contents below are printed:
6.
1234567890
Printer
abcdefghijklmnopqrstuvwxyz
ABC D E FG H IJ K LM N O PQ R STUVWXYZ
7.
Check whether the internal electronic simulator is working
Inter Simulator
properly.
Calibrate the touch screen (Touch the symbol on the screen,
five symbols appear in succession). It should be verified after
8.
the completion of the calibration test (Touch the symbol on
Screen Calibration
the screen again, five symbols appear in succession).
9.
Export the original data stored in the analyzer to *.txt format.
Export Data
10.
Reset the system password to the factory one.
Reset Password
11.
Restore the system setup to the factory setup.
Factory Setup
12.
Online update the analog board.
Analog Update
13.
Scan the system ID bar code.
System ID
14.
View the logs about the software running.
System Log
15.
3.
Delete all the data stored in the analyzer. Perform this action
after scanning the system ID bar code before shipment.
Clear Data
Press Return to go back to the System Setup screen.
4.5 External Simulator Test
Follow the procedures below to perform an external simulator test:
1.
Press the On/Off button on the left hand side of the analyzer to turn it on.
2.
Enter the user name and password manually, and then press
To scan the bar code of the user name, press
- 14 -
.
first, and then scan the bar code.

22.

i15 Blood Gas and Chemistry Analysis System Service Manual
Functional Checks
Figure 4-2 Enter User Name and Password
3. On the Main screen, press
to go to the Quality Control screen, and then press
.
4. Gently insert the external electronic simulator into the cartridge port, and press down to
ensure it clicks into place. If the electronic simulator is inserted properly, the system will
automatically perform a simulator test.
Figure 4-3 Insert External Simulator
NOTE:
Avoid touching the contact pads.
Never touch the simulator after inserting it into the system.
5. The system automatically starts to perform an external simulator test.
- 15 -

23.

i15 Blood Gas and Chemistry Analysis System Service Manual
Functional Checks
Figure 4-4 Measurement in Progress
6. The system automatically displays the test result upon the completion of the test.
Figure 4-5 External Simulator Test Result
7. View the result.
NOTE:
Upon the completion of the test, the message “Please Remove Cartridge” will be
displayed on the status bar at the bottom of the screen, and the simulator will be
ejected.
If the system passes the external simulator test, it can be used to analyze samples.
If the system fails the external simulator test, re-perform the test after the simulator
has been cleaned and completely dried or use another simulator to re-perform the
- 16 -

24.

i15 Blood Gas and Chemistry Analysis System Service Manual
Functional Checks
test. If the external simulator test succeeds, the analyzer can be used to run samples.
If the external simulator test still fails, run an internal simulator test. If the internal
simulator test succeeds, perform the external simulator test after the POGO PIN has
been cleaned and completely dried. If the external simulator test fails, the cam needs
to be replaced. If the external simulator test succeeds, the analyzer can be used to
run samples. If the internal simulator test fails, check the wire between the cam and
the analog board. If the wire connects well, run the internal simulator test again. If the
internal simulator test still fails, replace the cam.
8. Press Home to go to the Main screen.
9. Remove the external simulator from the system, and put it back.
NOTE:
Never remove the external simulator until the test is complete.
4.6 Control Test
Follow the steps below to perform a control test:
1.
Examine the package label of controls to ensure they have not expired.
2.
Remove an ampoule from the box of controls and equilibrate it to room temperature.
If oxygen is to be measured, the ampoule needs to stand at room temperature for at least 4
hours. If not, the ampoule needs to stand at room temperature for 30 minutes.
3.
Press the On/Off button on the left hand side of the analyzer to turn it on.
4.
Enter the user name and password manually, and then press
.
To enter the user name with the bar code scanner, press
first, and then scan the user
name bar code.
- 17 -

25.

i15 Blood Gas and Chemistry Analysis System Service Manual
Functional Checks
Figure 4-6 Enter User Name and Password
5.
On the Main screen, press
6.
Select the desired control type.
7.
to go to the Quality Control screen.
Press
to perform a control test for blood gases and blood chemistries.
Press
to perform a control test for Hct.
Press Scan Barcode, and scan the bar code on a new cartridge foil pouch.
If the bar code is scanned successfully, the system will beep and the scanner will be
automatically turned off. If the scanned data is valid, the system will display the screen for
the next procedure. If the scanned data is invalid, a message will pop up to prompt you.
Figure 4-7 Scan Bar Code
8.
Open the foil pouch and remove the cartridge from it.
- 18 -

26.

i15 Blood Gas and Chemistry Analysis System Service Manual
Functional Checks
NOTE:
Avoid tearing the bar code on the foil pouch.
For sample introduction with a capillary tube or an ampoule, insert a capillary
adaptor/ampoule adaptor into the fillport after removing the cartridge.
When inserting an ampoule adaptor into the fillport of a test cartridge, make sure the
indent of the adaptor is on the upside as shown in the picture below:
9.
Press Scan Barcode, and scan the bar code on the controls user manual.
If the bar code is scanned successfully, the system will beep and the scanner will be
automatically turned off. If the scanned data is valid, the system will display the screen for
the next procedure. If the scanned data is invalid, a message will pop up to prompt you.
Figure 4-8 Scan Bar Code
- 19 -

27.

i15 Blood Gas and Chemistry Analysis System Service Manual
Functional Checks
NOTE:
Keep the controls user manual for a future control test use.
10. Mix the controls completely by shaking the ampoule gently, and then tap the tip of the
ampoule carefully with your fingernail to remove any solution.
Figure 4-9 Mix Controls and Insert Cartridge
NOTE:
Avoid heating the ampoule with your hands when shaking the ampoule.
11. Open the ampoule by snapping off the top and immediately transfer control solution by
slowly drawing an appropriate amount of solution with a syringe or capillary tube from the
bottom of the ampoule.
NOTE:
When using an ampoule for sample introduction, you need not transfer control
solution. Insert the ampoule into the adaptor after opening it, and directly go to step
13.
Take protective measures when opening the ampoule, such as using gloves, tissue,
etc.
12. Insert the syringe or capillary tube into the fillport of the cartridge.
NOTE:
When using a syringe, discard the first 2 drops of solution first, then remove the
needle from it, and finally insert it into the fillport.
- 20 -

28.

i15 Blood Gas and Chemistry Analysis System Service Manual
Functional Checks
When using a capillary tube, directly insert the capillary tube into the adaptor till the
tube reaches the interface between the adaptor and the cartridge.
To avoid inaccurate test results, make sure there are no bubbles in the sample. If
bubbles continually exist, use a new ampoule and syringe or capillary tube to collect
samples again.
13. Gently insert the cartridge into the cartridge port, and carefully press down to ensure that it
clicks into place.
For a valid cartridge, the indicator in the cartridge port will turn green, and the system will
automatically aspirate calibrant. For an invalid cartridge, the indicator will turn red, the
cartridge will be ejected and a message will pop up to prompt you.
NOTE:
The cartridge can not be removed from the analyzer until the measurement is
complete.
Never inject the sample. It will be aspirated automatically.
14. The system automatically aspirates calibrant.
Figure 4-10 Aspirating Calibrant…
15. The system automatically performs calibration.
- 21 -

29.

i15 Blood Gas and Chemistry Analysis System Service Manual
Functional Checks
Figure 4-11 Calibration in Progress
16. The system automatically aspirates samples when calibration is complete.
Figure 4-12 Sampling…
17. The system automatically analyzes the sample upon the completion of sampling.
Figure 4-13 Measurement in Progress
- 22 -

30.

i15 Blood Gas and Chemistry Analysis System Service Manual
Functional Checks
18. The system automatically displays the test results upon the completion of the test.
Figure 4-14 Control Test Results
19. View the results.
NOTE:
The system will indicate whether the results are within or outside the acceptable
ranges with Under Control/Out of Control.
If a parameter fails calibration, the system will not be able to determine whether it is
under control, and will display Calibration Failure to prompt you.
The system will not report the result for a parameter failing the control test in the
patient sample analysis, if QC Lockout function is enabled in Setup. To report the
result for the parameter, repeat the control test till the parameter passes it.
If the results are outside the acceptable ranges, check the following items first, and
then perform another test.
Refer to the user manual to confirm that the test procedures are correct.
Test cartridges and controls are stored properly and have not expired.
The system passes the electronic simulator test.
If all the above items are verified, but the results are still outside the acceptable
ranges, please return the analyzer to EDAN for service.
20. Remove the test cartridge from the system.
21. Press Print to print the results.
22. Press Home to return to the Main screen.
- 23 -

31.

i15 Blood Gas and Chemistry Analysis System Service Manual
Troubleshooting
5 Maintenance
5.1 Maintenance Inspection
(1) Visual Inspection
Before using the system, the user should:
Inspect the analyzer for signs of damage.
Check the external cables for splits, cracks or signs of twisting.
Do not use the system if any damage occurs until the system is repaired by the professional
service personnel of EDAN or the authorized distributors.
(2) Routine Inspection
The following safety checks should be performed at least once every 24 months by a
qualified person who has adequate training, knowledge, and practical experience to perform
these tests.
Inspect the analyzer and accessories for mechanical and functional damage.
Inspect the safety related labels for legibility.
Verify the analyzer functions properly as described in the instructions for use.
If the analyzer is not functioning properly or fails any of the above tests, it has to be
repaired.
WARNING
Failure on the part of the responsible individual institution employing this equipment to
implement a satisfactory maintenance schedule may cause undue equipment failures.
(3) Mechanical Inspection
Make sure all exposed screws are tightly fixed.
5.2 Maintenance of the Analyzer
Keep the surface of the analyzer clean, free of dust and dirt.
Prevent any liquid from seeping into the analyzer.
Avoid shaking the analyzer violently when moving it to another place.
Do not scratch or damage the screen.
The gathering of dew on the screen may occur with abrupt temperature or humidity changes.
It is recommended that the system should be used in a stable environment.
- 24 -

32.

i15 Blood Gas and Chemistry Analysis System Service Manual
Troubleshooting
6 Principle Introduction
The system utilizes potentiometry and amperometry to determine the concentrations of blood
gases and blood chemistries, and utilizes conductivity to determine the concentration of Hct.
Many parameters can also be calculated such as cH+, HCO3-act, HCO3-std, BE(ecf), BE(B),
BB(B), and so on.
6.1 Measurement Method
Methods
Measurements are performed on undiluted specimens. Undiluted methods are also called direct
methods, whereas methods that dilute the sample are called indirect methods.
For electrolytes, direct methods measure the free ion concentration of analyte (apparent or free
ion activity) per unit volume of plasma water, and indirect methods measure the concentration of
analyte per unit volume of plasma. It is known that the direct method gives the clinically
significant result for electrolytes. When there is disagreement between the methods, such as when
the patient has abnormal total protein or lipid levels, it results from interference on the indirect
method. At normal levels of protein and lipids the systematic offset between methods is often
corrected for in commercial direct measuring instruments so that the normal ranges for all
instruments are in agreement. Sensors have been calibrated in the manufacture so that normal
ranges are in agreement with indirect reference methods at normal levels of total protein and
lipids.
Direct measurement of hematocrit by the conductometric technique gives a result related to the
non-conducting excluded volume fraction of the sample. Red blood cell volume is the
predominant component of the nonconducting volume, but proteins, lipids, and white blood cells
also contribute. Elevated hematocrit readings are expected at abnormally elevated levels of these
components. Decreased hematocrit readings are expected at abnormally low levels of protein,
such as found in hemodiluted samples taken from cardiopulmonary bypass. Osmotic imbalance
causes a discrepancy between direct (conductometric, spun) and indirect (Coulter) measurements
because of variation in the meancell volume.
Sensors
Sensor refers to the electrode embedded in test cartridges. There are three different types of
sensors:
- 25 -

33.

i15 Blood Gas and Chemistry Analysis System Service Manual
Potentiometric Sensor
Amperometric Sensor
Conductometric Sensor
Troubleshooting
Potentiometry: A potential is recorded using a voltmeter, which relates to the concentration of
the sample. A reference electrode is used to provide a stable, fixed potential against which other
potential differences can be measured. This measurement technique is used for pH, pCO2 and
electrolytes.
Amperometry: The magnitude of an electrical flow of current is proportional to the
concentration of the substance being oxidized or reduced at an electrode. This measurement
technique is used for pO2, glucose and lactic acid.
Conductivity: The specific impedance of a sample as measured by two conducting electrodes
held at a constant voltage is directly proportional to the conductive properties of the sample. This
technique is used for Hct.
6.2 Determination of Test Results
6.2.1 Determination of the Analyte Concentration
The concentration of the analyte is determined with the potentiometric and amperometric sensors.
For both sensors, the analyte concentration is calculated using:
The known analyte concentration of calibrant solution;
The measured voltage or current of the calibrant solution.
The measured voltage or current of the sample.
For potentiometric sensors, the analyte activity in the sample is calculated from the Nerst
equation:
Esample – Ecalibrant = S log (αsample/αcalibrant)
Where E denotes the potential, α denotes the activity of an ion, and S denotes the slope of the
sensor.
- 26 -

34.

i15 Blood Gas and Chemistry Analysis System Service Manual
Troubleshooting
6.2.2 Determination of Cell Concentration
Hct
In whole blood, the cellular constituents, red and white blood cells and platelets do not conduct
electricity, while plasma does. For a sample whose electrolyte concentration is given, the more
cells, the less conductivity. The total cell concentration of the whole blood is determined from:
The known value of the electrolyte concentration in the calibrant.
The measured electrolyte concentration of the sample.
The measured conductivity signal generated by the calibrant.
The measured conductivity signal generated by the sample.
CPB
When samples with abnormally low protein levels are tested, the system need to use the CPB
compensation algorithm. The CPB compensation algorithm is specially intended for use when
samples are taken from patients on cardiopulmonary bypass. It also applies to the adult whose
protein levels are abnormally low.
6.3 Equations for Calculated Parameters
cH+
Hydrogen ion concentration
cH+ = 10 (9 – pH) [nmol/L]
HCO3-act
Bicarbonate ion concentration
HCO3-act = 0.0307 ×pCO2 × 10(pH - 6.105) [mmol/L]
HCO3-std
Bicarbonate ion concentration normalized to a pCO2 40mmHg
HCO3-std = 24.5 + 0.9 × A + [(A - 2.9)2 × (2.65 + 0.31 × tHb (est))] / 1000 [mmol/L]
Where A = BE (B) – [0.2 × tHb (est) × (100 - sO2 (est))] / 100, tHb (est) can be entered by users,
and the default value for tHb (est) is 15g/dL.
- 27 -

35.

i15 Blood Gas and Chemistry Analysis System Service Manual
Troubleshooting
BE (ecf)
Base excess (ecf)
BE (ecf) = HCO3-act - 24.8 + (16.2 × (pH - 7.40)) [mmol/L]
BE (B)
Base excess (B)
BE (B) = (1 - 0.014 × tHb (est)) × [HCO3-act - 24.8 + (1.43 × tHb (est) + 7.7) × (pH - 7.40)]
[mmol/L]
Where tHb (est) can be entered by users, and the default value for tHb (est) is 15g/dL.
BB (B)
Buffer Base
BB (B) = BE (B) + 41.7 + 0.42 × tHb (est) [mmol/L]
Where tHb (est) can be entered by users, and the default value for tHb (est) is 15g/dL.
ctCO2
Total carbon dioxide
ctCO2 = HCO3-act + 0.0307 × pCO2 [mmol/L]
Ca++ (7.4)
The ionized calcium concentration of blood normalized to pH 7.4
Ca++ (7.4) = Ca++ × 10 [0.178 ×(pH - 7.40)] [mmol/L]
AnGap
An approximation of the difference between measured cations and measured anions in the
sample
AnGap = (Na+ + K+) - (Cl- + HCO3-act) [mmol/L]
tHb (est)
An estimation of the hemoglobin contained in the sample
tHb (est) = MCHC × Hct / 100 [g/dL]
- 28 -

36.

i15 Blood Gas and Chemistry Analysis System Service Manual
Troubleshooting
Where MCHC is the mean corpuscular hemoglobin concentration, and can be entered by users.
The default value for MCHC is 34g/dL. The unit for Hct is %PCV.
sO2 (est)
An estimation of hemoglobin oxygen saturation: a ratio of hemoglobin bound to oxygen to the
total amount of hemoglobin able to bind oxygen
sO2 (est )
pO2*3 pO2*
100
pO2*3 pO2*
[%]
pO2 (A-a)
Alveolar-arterial oxygen tension difference
pO2 (A-a) = pO2 (A) - pO2 (a) [mmHg]
pO2 (a/A)
Arterial-alveolar oxygen tension ratio
pO2 (a/A) = pO2 (a) / pO2 (A) [mmHg]
RI
Respiratory index: the ratio of the alveolar-arterial blood oxygen-pressure difference to arterial
pO2
RI = pO2 (A-a) / pO2 (a)
pO2 / FIO2
The ratio of arterial pO2 to the fraction of inspired oxygen
pO2 / FIO2 = pO2 / FIO2 [mmHg]
pH (T)
pH value corrected for entered patient temperature
pH (T) = pH – [0.0147 + 0.0065 × (pH – 7.4)] (T – 37)
cH+ (T)
Hydrogen ion concentration corrected for entered patient temperature
- 29 -

37.

i15 Blood Gas and Chemistry Analysis System Service Manual
Troubleshooting
cH+ (T) = 10 (9 – pH (T)) [nmol/L]
pCO2 (T)
pCO2 corrected for entered patient temperature
pCO2 (T) = pCO2 × 100.019(T - 37) [mmHg]
pO2 (T)
pO2 corrected for entered patient temperature
[mmHg]
pO2 (A-a) (T)
Alveolar-arterial oxygen tension difference corrected for entered patient temperature
pO2 (A-a) (T) = pO2 (A) (T) - pO2 (a) (T) [mmHg]
pO2 (a/A) (T)
Arterial-alveolar oxygen tension ratio corrected for entered patient temperature
pO2 (a/A) (T) = pO2 (a) (T) / pO2 (A) (T) [mmHg]
RI (T)
Respiratory index: the ratio of the alveolar-arterial blood oxygen-pressure difference to arterial
pO2 when both values are corrected for patient temperature
RI (T) = pO2 (A-a) (T) / pO2 (a) (T)
pO2 (T)/FIO2
The ratio of arterial pO2 to the fraction of inspired oxygen corrected for the entered patient
temperature
pO2 (T) / FIO2 = pO2 (T) / FIO2 [mmHg]
mOsm
mOsm is generally determined by Na+, K+, Glu and Bun.
- 30 -

38.

i15 Blood Gas and Chemistry Analysis System Service Manual
Troubleshooting
mOsm = 1.86 × (Na+ + K+) +Glu + BUN +9
In the equation:
Glu is the Glu test result by performing test with Blood Gas and Chemistry Test Cartridge on i15
Blood Gas and Chemistry Analyzer and the unit for Glu test result is mmol/L. If Glu can not be
measured directly, its default value for Glul is 5mmol/L.
BUN is the BUN test result by performing test with Blood Gas and Chemistry Test Cartridge on
i15 Blood Gas and Chemistry Analyzer and the unit for BUN test result is mmol/L. If BUN can
not be measured directly, its default value for BUN is 3mmol/L.
6.4 System Principle Block Diagram
On the left side of the analyzer, there are 6 interfaces:
RJ45 Interface (Network)
DB9 Interface (RS232)
4 USB Interfaces
6.4.1 RJ45 Interface
The analyzer is configured with a RJ45 interface. It is used to connect the analyzer to a computer
for updating or to a DMS.
6.4.2 DB9 Interface
The analyzer is configured with a DB9 interface. It allows engineers from the manufacture and
service personnel to troubleshoot.
6.4.3 USB Interface
The analyzer is configured with 4 USB interfaces. It is used to connect external devices
supporting USB protocol to the analyzer, such as a USB drive, a mouth, a printer, a scanner.
NOTE:
Only the following external scanners should be connected to the analyzer through
USB ports: Honeywell 1900.
Only the following external printers should be connected to the analyzer through
USB ports: HP LaserJet P401DN, and HP LaserJet P1606DN.
- 31 -

39.

i15 Blood Gas and Chemistry Analysis System Service Manual
Troubleshooting
7 Troubleshooting
EDAN supports replacement of PCB and major subassemblies for the analyzer. When
replacement is necessary, follow the procedures described in chapter 9 Disassembling the
Analyzer.
7.1 Booting Failures
Phenomenon
Possible Cause
Solution
Check whether the power cord is
damaged, and whether the power
adaptor is damaged. If so, replace
No
AC
mains
are them. Check whether the power
connected, and the battery cord
is empty.
and
power
adaptor
are
correctly connected to the power
socket of the analyzer and the
After switching on, LCD has
hospital AC outlet. If not, connect
no
them correctly.
display;
the
power
indicator is off; the exhaust
Check
fan does not run.
connecting the power button are
The connecting wires are loose,
loose.
whether
and
the
whether
the
wires
wires
connecting the upper assembly and
lower assembly are loose. If so,
connect them again.
Power board is defective.
Main
control
board
Replace the power board.
is
defective.
Replace the main control board.
The analyzer is struck by
strong high voltage, such as Check the power supply.
Abrupt switching off.
lightening strike.
Bad
power
performance
- 32 -
supply
Replace power supply.

40.

i15 Blood Gas and Chemistry Analysis System Service Manual
Troubleshooting
Check whether the power cord and
power
adaptor
connect
well,
whether the power adaptor and the
Bad connection
power socket connect well, and
whether the power cord and the
Abrupt switching off.
electrical outlet connect well. If not,
connect them well.
Main
control
board
is
defective.
Replace the main control board.
7.2 Display Failures
Phenomenon
Possible Cause
Solution
Check
whether
the
wires
the The connecting wires are connecting the LCD are loose, and
whether the wires connecting the
power indicator is on, and loose.
upper assembly and lower assembly
the exhaust fan works
After
switching
on,
properly, but LCD has no
are loose. If so, connect them again.
display,
Replace the LCD.
or
display
is LCD is defective.
abnormal or dim.
Main
control
board
is
defective.
Replace the main control board.
7.3 Touch Screen Failures
Phenomenon
Possible Cause
The connecting wires are
loose.
Solution
Check
whether
the
connecting
wires are loose. If so, connect them
again.
The touch screen does not
response after pressing it. The work area of the touch Adjust the silicon loop so that the
But it responses when screen is covered by the work area of the touch screen is not
operating the system with a silicon loop.
mouse.
covered.
Touch screen drive board is Replace the touch screen drive
defective.
board.
Touch screen is defective.
Replace the touch screen.
- 33 -

41.

i15 Blood Gas and Chemistry Analysis System Service Manual
Main
control
Troubleshooting
board
is
defective.
The
touch
position
is The touch screen is not
ineffective.
calibrated.
Replace the main control board.
Calibrate
the
according
to
touch
the
screen
section
4.5
Diagnostics.
7.4 Operation Failures
Phenomenon
Possible Cause
The
test
cartridge
Solution
is
inserted into the system
prior to a test.
A test cartridge can not be
ejected automatically.
The system is shut down
during a test.
object, and remove the test cartridge
Reboot the system. After about 1
minute, the test cartridge will be
ejected automatically.
a test.
prompt
pull the middle hole with a pointed
from the system.
The system is dead during
The
Remove the test cartridge ejector cap,
Reboot the system. After about 1
minute, the test cartridge will be
ejected automatically.
“Ambient The ambient temperature Use the system in an environment
Temperature Out of Range”
is not within 10°C - with the ambient temperature within
appears.
31°C.
10°C - 31°C.
Forget
The prompt “Sample is Not
Aspirated Properly” appears.
to
connect
a
syringe/capillary tube to
the fillport.
The sample is not enough.
The prompt “Calibrant Pack
Improper Removal. Please The calibrant fluid pack is
Replace
Calibrant
Pack” removed improperly.
prompt
Expired” appears.
“Cartridge The
date
setup
incorrect.
- 34 -
the fillport.
Perform
the
test
with
enough
samples.
Replace the calibrant fluid pack
following
the
user
manual
instructions.
appears.
The
Connect a syringe/capillary tube to
is
Set the date to be the current date.

42.

i15 Blood Gas and Chemistry Analysis System Service Manual
Troubleshooting
Use a new test cartridge that has not
The test cartridge expired.
The
date
setup
expired.
is
Set the date to be the current date.
The prompt “Calibrant Pack incorrect.
Expired” appears.
The calibrant fluid pack Use a new calibrant fluid pack that
expired.
has not expired.
7.5 Printer Failures
Phenomenon
Possible Cause
Solution
The connecting wires are
There is no response after loose.
pressing Print.
The printout is not clear.
whether
the
connecting
wires are loose. If so, connect them
again.
The printer is defective.
Paper jam or slant
Check
Replace the printer.
The printer casing was not Open the printer casing, and shut
shut properly.
The
the casing again.
printer
head
is
defective.
Replace the printer head.
7.6 Network Failures
Phenomenon
Possible Cause
The
network
wire
Solution
is
damaged.
Replace the network wire.
Set the network according to the
The analyzer can not be Network setup error
user manual.
connected to a network.
Main
control
board
defective.
Power board is defective.
Network setup error
Wi-Fi connection failure
Wi-Fi board is defective.
- 35 -
is
Replace the main control board.
Replace the power board.
Set the network according to the
user manual.
Replace the Wi-Fi board.

43.

i15 Blood Gas and Chemistry Analysis System Service Manual
Main
control
board
defective.
Troubleshooting
is
Replace the main control board.
7.7 Diagnosis Database
NOTE:
Only service engineers and engineers from the manufacturer can enter the diagnosis
database.
Diagnosis data help to troubleshoot. The diagnosis database displays diagnosis data for the most
recent month by default, and it can store up to 500 data entries. The following operations can be
performed in the diagnosis database: exporting diagnosis data to a removable disk (such as a
USB drive), viewing the details of diagnosis data, searching for and printing diagnosis data, etc.
7.7.1 Searching for Diagnosis Data
1.
On the Database screen, press
to open the diagnosis database.
Figure 7-1 Diagnosis Database Screen
2.
Press Search.
- 36 -

44.

i15 Blood Gas and Chemistry Analysis System Service Manual
Figure 7-2 Enter Search Conditions
3.
Enter the search conditions, and press OK.
4.
The system displays the results.
Figure 7-3 Search Results
5.
Press Return to go back to the Database screen.
7.7.2 Viewing Details of Diagnosis Data
1.
Select the desired diagnosis data.
2.
Press Details.
3.
The system displays the details.
- 37 -
Troubleshooting

45.

i15 Blood Gas and Chemistry Analysis System Service Manual
Troubleshooting
Figure 7-4 Details of Diagnosis Data
4.
View the details.
5.
Press Return to go back to the Database screen.
7.7.3 Exporting/Printing Diagnosis Data
1.
On the Database screen, press
2.
Select the desired diagnosis data.
to open the diagnosis database.
To
Do this
Export
Insert a removable disk into the analyzer, and then press Export.
Print
Press Print.
NOTE:
If no diagnosis data is selected before pressing Export/Print, all of the data stored in the
diagnosis database will be exported / printed.
Here is an example of diagnosis report:
System ID
201303200028
Report Type
Diagnosis
Print Time
2012-03-20 15:20:00
Test Time
2012-03-20 15:20:00
Info. Code
2066
- 38 -

46.

i15 Blood Gas and Chemistry Analysis System Service Manual
Module’s Malfunction Verification
8 Module’s Malfunction Verification
When a module is suspected of malfunction, verify it with the methods described in this chapter.
NOTE:
For all the interfaces, the first pin is the one with a square solder pad.
8.1 Verification of Main Control Board’ Circuit Failure
Follow the steps below to verify the main control board:
1) Disassemble the analyzer following the procedures described in section 9.2.
2) Turn on the analyzer (Powered by AC or battery).
3) Measure the voltage to earth of the pins listed below with a multimeter:
Item
Pin
Reference Range
1.
TP8
1.71~1.89V
2.
TP9
3.13~3.47V
3.
TP12
0.95~1.05V
4.
TP13
0.95~1.05V
5.
TP14
2.37~2.63V
6.
TP49
3.13~3.47V
- 39 -

47.

i15 Blood Gas and Chemistry Analysis System Service Manual
Module’s Malfunction Verification
4) Compare the measurement results with the reference ranges in the list. If there is a result
exceeding its reference range, it is confirmed that the main control board malfunctions, and
it is recommended to replace the main control board.
8.2 Verification of the Power Board’ Circuit Failure
Follow the steps below to verify the power board:
1) Disassemble the analyzer following the procedures described in section 9.2.
2) Remove the upper assembly.
3) Turn on the analyzer (Powered by AC or battery).
4) Measure the voltage to earth of the pins listed below with a multimeter.
Item
Pin
Reference Range
1.
TP12
11.40~12.60V
2.
TP13
4.75~5.25V
3.
TP9
4.75~5.25V
- 40 -

48.

i15 Blood Gas and Chemistry Analysis System Service Manual
Module’s Malfunction Verification
5) Compare the measurement results with the reference ranges in the list. If there is a result
exceeding its reference range, it is confirmed that the power board malfunctions, and it is
recommended to replace the power board.
8.4 Verifying Malfunction of the Touch Screen Drive Board
Follow the steps below to verity the touch screen drive board:
1) Disassemble the analyzer following the procedures described in section 9.2.
2) Remove the upper assembly.
3) Turn on the analyzer (Powered by AC or battery).
4) Measure the voltage to earth of the pins listed below with a multimeter.
Item
Pin
Reference Result
1.
TP<+3V3>
3.3V±0.165V
- 41 -

49.

i15 Blood Gas and Chemistry Analysis System Service Manual
Module’s Malfunction Verification
5) Compare the measurement results with the reference ranges in the list. If there is a result
exceeding its reference range, it is confirmed that the touch screen drive board malfunctions,
and it is recommended to replace the touch screen drive board.
8.5 Verifying Malfunction of the Scanner
Follow the steps below to verify the bar code scanner:
1) Disassemble the analyzer following the procedures described in section 9.2.
2) Remove the bar code scanner.
3) Connect the bar code scanner to a computer through the USB port.
4) Operate the bar code scanner with the computer.
5) If you fail to operate the bar code scanner, it is confirmed that the bar code scanner
malfunctions, and it is recommended to replace the bar code scanner.
NOTE:
It is necessary that the SNAPI Sample Application software has been loaded on the
computer before connecting the bar code scanner to the computer. If the software has
been loaded, the symbol
will appear.
- 42 -

50.

i15 Blood Gas and Chemistry Analysis System Service Manual
Disassembling the Analyzer
9 Disassembling the Analyzer
WARNING
1. Only qualified service personnel should open the analyzer housing.
2. Switch off the analyzer and remove the power supply before disassembling the
analyzer.
3. After any repair of the analyzer, perform functional checks prior to use.
9.1 Tools Required
1. A cross-head screwdriver
2. A M3 nut driver
9.2 Disassembling the Main Unit
The main unit consists of two parts: upper assembly and lower assembly.
Follow the steps below to disassemble the main unit:
1. Carefully place the analyzer upside down on a flat surface covered with cloth or other
protecting pad.
2. With a cross-head screwdriver, remove the 7 screws securing the upper assembly to the lower
assembly.
3. Hold both upper and lower assembly simultaneously, and place the analyzer upward again.
4. Unlock the calibrant fluid pack chamber lock with the key, and then pull the calibrant fluid
pack ejector to open the calibrant fluid pack chamber door.
5. Slightly lift up the upper assembly, and disconnect the connecting wire on the shrouded &
ejector IDC header to separate the upper assembly and the lower assembly.
6. Carefully lift up the upper assembly, and place it upside down on the left side of the lower
assembly.
7. Disconnect the printer cable to earth to disassemble the upper and lower assembly.
- 43 -

51.

i15 Blood Gas and Chemistry Analysis System Service Manual
Disassembling the Analyzer
3
1
5
2
4
1 – Upper assembly
2 – Upper assembly & lower assembly connecting wire
3 – Printer cable to earth
4 – Shrouded & ejector IDC header
5 – Lower assembly
9.3 Disassembling the Lower Assembly
9.3.1 Replacing the Main Control Board
Follow the steps below to remove the main control board:
1. Separate the upper assembly and the lower assembly.
2. Disconnect all the wires on the main control board.
3. Remove the 5 screws securing the main control board.
4. Hold the left side of the main control board with your hand, sway it back and forth, and
remove the main control board carefully.
- 44 -

52.

i15 Blood Gas and Chemistry Analysis System Service Manual
Disassembling the Analyzer
To install a main control board:
Install a main control board in the reverse order. Refer to the following illustration when
connecting wires, and ensure they are all correctly connected.
1
3
2
No.
Connect to
1
Upper assembly
2
Analog communication board
3
Speaker
4
Bar code scanner
5
Wi-Fi board
9.3.2 Replacing the Wi-Fi Board
Follow the steps below to remove the Wi-Fi board:
1. Separate the upper assembly and the lower assembly.
2. Disconnect the wire on the Wi-Fi board.
3. Remove the Wi-Fi antenna socket.
4. Remove the 2 screws securing the Wi-Fi board.
5. Remove the Wi-Fi board carefully.
- 45 -
4
4
5

53.

i15 Blood Gas and Chemistry Analysis System Service Manual
To install a Wi-Fi board:
Install a Wi-Fi board in the reverse order.
9.3.3 Replacing the Chamber Door Detection Board
Follow the steps below to remove the chamber door detection board:
1. Separate the upper assembly and the lower assembly.
2. Disconnect the wire on the chamber door detection board.
3. Remove the screw securing the chamber door detection board.
4. Remove the chamber door detection board carefully.
To install a chamber door detection board:
Install a chamber door detection board in the reverse order.
- 46 -
Disassembling the Analyzer

54.

i15 Blood Gas and Chemistry Analysis System Service Manual
Disassembling the Analyzer
9.3.4 Replacing the Ambient Temperature Detection Board
Follow the steps below to remove the ambient temperature detection board:
1. Separate the upper assembly and the lower assembly.
2. Disconnect the wire on the ambient temperature detection board.
3. Remove the screw securing the ambient temperature detection board.
4. Remove the ambient temperature detection board carefully.
To install an ambient temperature detection board:
Install an ambient temperature detection board in the reverse order.
9.3.5 Replacing the Analog Board
Operation Procedures
NOTE
Remove the screws located
at the left picture.
- 47 -

55.

i15 Blood Gas and Chemistry Analysis System Service Manual
Remove
the
connection
cable between the main
control board and analog
board as the left picture
shows.
Remove
the
connection
cable between the main
control
board
and
the
trumpet as the left picture
shows.
Open the cover of the FPC
connector which connect
the main control board and
barcode module.
- 48 -
Disassembling the Analyzer

56.

i15 Blood Gas and Chemistry Analysis System Service Manual
Open the cover of the FPC
connector which connect
the main control board and
WIFI module.
Lift the main control board
upward.
- 49 -
Disassembling the Analyzer

57.

i15 Blood Gas and Chemistry Analysis System Service Manual
Remove
the
connection
cable of chamber detection
board.
Remove the FPC cable
between the analog board
and the cam.
- 50 -
Disassembling the Analyzer

58.

i15 Blood Gas and Chemistry Analysis System Service Manual
Disassembling the Analyzer
Remove the copper pillars
from the board with the
Copper pillar
help of some tools. Then
the board can be separated.
Reassemble a new analog
board as the reverse
procedures.
9.3.6 Replacing the Cam Machine
Steps
NOTE
Remove the seven screws as the
picture shows with a cross-head
screwdriver.
Dismantle the
upper
assembly
- 51 -

59.

i15 Blood Gas and Chemistry Analysis System Service Manual
Lift the upper assembly for 10cm.
Disconnect the cable from the socket
located at the lower assembly.
Disconnect
the
connection
cable
which connects the upper and lower
cover.
Separate the upper and lower cover
and then pull out the ground wire.
- 52 -
Disassembling the Analyzer

60.

i15 Blood Gas and Chemistry Analysis System Service Manual
Open the black cover of the FPC
connector.
Gently remove the FPC cable.
Dismantle the
Pull out the serial cable and trumpet
main control
wire.
board (PCBA)
Remove the five screws which are
used to fix the main control board
(PCBA).
Gently lift the main control board
(PCBA) upward.
Remove the screw which is used to fix
the detection board and then the
detection board can be separated from
Remove
Cam the module.
Module
- 53 -
Disassembling the Analyzer

61.

i15 Blood Gas and Chemistry Analysis System Service Manual
Remove the four screws which are
used to fix the cam and then the cam
be removed.
Open the black cover of the FPC
connector and then remove the FPC
cable from the connector.
Hold the cam connection cable tightly
and then disconnect it from the cable.
Connect the FPC cable with a new
cam and then close the black cover of
the connector.
Press down the cam connection cable
and fix the cam assembly with the
4-M3*8 screw.
Assemble
Cam
- 54 -
Disassembling the Analyzer

62.

i15 Blood Gas and Chemistry Analysis System Service Manual
Fix the temperature detection board
with 1-M3*8 screw.
Mount the main control board onto the
analog board (please pay attention to
the connection board between the
main control board and the analog
board).
Assemble
main
control
board.
Tightly secure the analog board and
the main control board with 5-M3*6
screws screwed by torque (3.0 4.0
kg.cm)
Connect serial cable, trumpet wire,
FPC cables.
Connect the grounding cover
Assemble
upper and
lower cover
- 55 -
Disassembling the Analyzer

63.

i15 Blood Gas and Chemistry Analysis System Service Manual
Disassembling the Analyzer
Connect the connection cable which
connects the upper and lower cover.
Secure the screws.
9.3.7 Replacing the Power Board
Follow the steps below to remove the power board:
1. Separate the upper assembly and the lower assembly.
2. Remove the main control board and the cam machine.
3. Disconnect all the wires on the power board.
4. Remove the 5 screws securing the power board.
5. Remove the two screws fixing the fan and then the fan can be removed.
6. Remove the power board carefully.
To install a power board:
Install a power board in the reverse order. Refer to the following illustration when connecting
wires, and ensure they are all correctly connected.
- 56 -

64.

i15 Blood Gas and Chemistry Analysis System Service Manual
Disassembling the Analyzer
2
1
4
3
No.
Connect to
1
Interface connection board
2
Exhaust fan
3
Battery
4
NTC thermal resistor
9.3.8 Replacing the Power Interface Board
Follow the steps below to remove the power interface board:
1. Separate the upper assembly and the lower assembly.
2. Remove the main control board and the analog board.
3. Disconnect all the wires on the power interface board.
4. Remove the 2 screws securing the power interface board.
5. Remove the power interface board carefully.
To install a power interface board:
Install a power interface board in the reverse order. Ensure the wire is correctly connected.
- 57 -

65.

i15 Blood Gas and Chemistry Analysis System Service Manual
Disassembling the Analyzer
9.3.9 Replacing the Exhaust Fan
Follow the steps below to remove the exhaust fan:
1. Separate the upper assembly and the lower assembly.
2. Disconnect the wire on the exhaust fan.
3. Remove the 2 screws securing the exhaust fan.
4. Remove the exhaust fan carefully.
To install an exhaust fan:
Install an exhaust fan in the reverse order. Ensure the wire is on the left side of the exhaust fan as
shown in the picture above.
- 58 -

66.

i15 Blood Gas and Chemistry Analysis System Service Manual
Disassembling the Analyzer
9.3.10 Replacing the Speaker
Follow the steps below to remove the speaker:
1. Separate the upper assembly and the lower assembly.
2. Disconnect the wire connecting the speaker on the main control board.
3. Remove the 2 screws securing the speaker.
4. Remove the speaker carefully.
To install a speaker:
Install a speaker in the reverse order. Ensure that the wire is correctly connected to the main
control board.
9.3.11 Replacing the Bar Code Scanner
Follow the steps below to remove the bar code scanner:
1. Separate the upper assembly and the lower assembly.
2. Disconnect the wires connecting the bar code scanner on the main control board.
3. Remove the 2 screws securing the bar code scanner.
4. Remove the bar code scanner carefully.
To install a bar code scanner:
Install a bar code scanner in the reverse order. Ensure that the wire is correctly connected to the
main control board.
- 59 -

67.

i15 Blood Gas and Chemistry Analysis System Service Manual
Disassembling the Analyzer
NOTE:
After installing a new bar code scanner, the bar code scanner should be set.
Follow the steps below to set a bar code scanner:
1. Press the power button on the left side of the analyzer to turn it on.
2. Press
in the User Login screen.
3. Scan the following bar codes in sequence:
- 60 -

68.

i15 Blood Gas and Chemistry Analysis System Service Manual
Disassembling the Analyzer
4. Enter the user name and password in the User Login screen.
5. Press Setup
Diagnostics
System ID. The bar code scanner will be turned on
automatically.
6. Scan the system ID bar code.
7. Check whether the last 12 digits of the scanned data are consistent with those of the system
ID. If they are consistent, the bar code scanner is successfully set.
9.3.12 Replacing the Spring Located in Test Cartridge Base
Operation Procedures
Note
- 61 -

69.

i15 Blood Gas and Chemistry Analysis System Service Manual
Remove the four screws as the right picture
shows.
Remove the two screws as the right picture
shows.
Disconnect the connection wire of vacuum
pump from the PCB board and remove the
vacuum pump from the slot as the right picture
shows.
- 62 -
Disassembling the Analyzer

70.

i15 Blood Gas and Chemistry Analysis System Service Manual
Disassembling the Analyzer
Remove the test cartridge base from its slot as
the right picture shows.
Lift the spring support upward.
Remove the spring from its support.
Insert a new spring into the support and
reassemble as the reverse procedures.
9.3.13 Cleaning and Replacing Vacuum Pump
Operation Procedures
If there is any need to replace the vacuum pump because of some damages, please replace the
vacuum pump as the following procedures:
Disconnect the tube from the vacuum pump
- 63 -

71.

i15 Blood Gas and Chemistry Analysis System Service Manual
Disassembling the Analyzer
Separate the vacuum pump from its motor
Remove the piston head by turning it anti-clockwise from the vacuum pump’s motor
Replace a new piston head and assemble it into the vacuum pump’s motor
Insert the vacuum pump’s motor into the vacuum pump.
Insert the tube into the connector of the vacuum pump and seal the joint with AB adhesive.
Assemble the vacuum pump as reverse procedures.
9.3.14 Replacing FPC Cable
Change FPC Cable
Operation Procedures
Note
- 64 -

72.

i15 Blood Gas and Chemistry Analysis System Service Manual
Disassemble
the Remove the five screws as
cam module
the right picture shows.
Remove the cam’s Remove the four screws as
cover
the right picture shows.
Remove the cover module.
- 65 -
Disassembling the Analyzer

73.

i15 Blood Gas and Chemistry Analysis System Service Manual
Remove the FPC
Smear some alcohol with
cable from the
cotton swab on the
heater
hot-melt adhesive which is
used to secure the FPC
connector.
Remove
the
holt-melt
adhesive
after
several
seconds
with
hands
or
some devices, and then
open the cover of the
connector
as
the
right
picture shows.
Remove the FPC cable
from the connector as the
right picture shows.
- 66 -
Disassembling the Analyzer

74.

i15 Blood Gas and Chemistry Analysis System Service Manual
Remove the motor
Remove the two screws as
the right picture shows
Disconnect the cable as the
right picture shows
- 67 -
Disassembling the Analyzer

75.

i15 Blood Gas and Chemistry Analysis System Service Manual
Remove the motor from
the cam as the right picture
shows
Remove the FPC Smear alcohol with the
cable
Cam
from
the cotton
swab
on
the
hot-melt adhesive located
which is used to secure the
FPC connector as the right
picture shows.
- 68 -
Disassembling the Analyzer

76.

i15 Blood Gas and Chemistry Analysis System Service Manual
Remove
the
holt-melt
adhesive
after
several
seconds as the right picture
shows.
Open the cover of the
connector and disconnect
the FPC cable from the
connector.
Disassemble front
As the right picture shows,
board assembly.
turn the turntable
anticlockwise to lower the
valve rod.
- 69 -
Disassembling the Analyzer

77.

i15 Blood Gas and Chemistry Analysis System Service Manual
Disassembling the Analyzer
As the right picture shows,
support up front board with
hands to remove the valve
rod from the turn table.
As the right picture shows,
remove the front board
from the cam assembly.
The front board is as the
right picture shows.
Disassemble
front board
the As the following picture shows, remove the two screws and separate the
cover from the front board.
- 70 -

78.

i15 Blood Gas and Chemistry Analysis System Service Manual
Disassembling the Analyzer
As the following picture shows, remove the four screws and separate the
cover from the front board.
Disassemble FPC As the right picture shows,
cable
from
front board.
the pull out the FPC cable
from the heater.
- 71 -

79.

i15 Blood Gas and Chemistry Analysis System Service Manual
Disassembling the Analyzer
As the right picture shows,
open the black cover of the
FPC cover.
As the right picture shows,
pull out the FPC cable.
As the right pictures, pull
back the FPC cable.
9.3.15 Replacing the Cable of Temperature Detection Board
Operation Procedures
NOTE
- 72 -

80.

i15 Blood Gas and Chemistry Analysis System Service Manual
Disassemble the cable of temperature detection
board.
Remove the screw as the right picture shows
and then remove the cover.
- 73 -
Disassembling the Analyzer

81.

i15 Blood Gas and Chemistry Analysis System Service Manual
Smear alcohol with the cotton swab on the
hot-melt adhesive and then remove the
adhesive after several seconds.
Open the cover of the FPC terminal and then
remove the FPC cable from the terminal.
- 74 -
Disassembling the Analyzer

82.

i15 Blood Gas and Chemistry Analysis System Service Manual
Disassembling the Analyzer
Remove the cable of the temperature detection
board
9.3.16 Replacing and Cleaning Needle
Clean L-shape Needle
1.
Check whether L-shape needle has been blocked.
1) Observe whether there are foreign objectives around the internal surface of the needle (in
many cases, the foreign objects will be accumulated in the point of the needle).
2) If the operator can obviously feel the obvious resistance when pulling or pushing the piston
of the syringe, the point of the needle has been blocked by the foreign objects.
2.
If the needle has been blocked, clean the needle as the following procedures:
1) Slowly pull the piston of vacuum pump to the maximum range.
2) Insert the 1mL syringe into the L-shaped needle and then move out the 1mL syringe from the
needle.
3) Quickly push the piston of the vacuum pump to blow out the foreign objects from the
needle.
- 75 -

83.

i15 Blood Gas and Chemistry Analysis System Service Manual
Disassembling the Analyzer
Change the test cartridge’s base assembly
If the L-shaped needle has been blocked heavily, the needle can be changed as the following
procedures:
1.
Disconnect the tubing from the test cartridge’s base assembly.
2.
Remove the spring from the test cartridge’s base assembly.
3.
Connect the tubing into the test cartridge’s base assembly, then apply some AB adhesive on
the joint to prevent air leakage.
Clean Calibrant Fluid Pack’s Needle
1.
Check whether there are foreign objects in the calibrant fluid pack’s needle. If yes, clean the
needle in accordance with Step 5 and Step 6.
2.
As Picture 12 shows, make sure the valve of the calibrant fluid pack is closed.
3.
Use the syringe to aspirate the air. If the piston of the syringe can be moved smoothly, it
proves the distance between the needle and the air valve has not been blocked.
4.
As Picture 13 shows, close air valve by pressing it with hands and then open the valve of
calibrant fluid pack.
5.
Use the syringe to aspirate calibration solutions. If the piston of the syringe can be moved
smoothly and the calibration solutions can be aspirated smoothly, it proves that the distance
between the valve of the calibrant fluid pack and the needle has not been blocked.
6.
As Picture 14 shows, insert 1mL or 2mL syringe into the needle. Quickly pull the piston
outward so that the foreign object can be aspirated out.
NOTE:
After the valve of the calibrant fluid pack has been opened, the calibration solution can only be
aspirated out, or it will be contaminated.
- 76 -

84.

i15 Blood Gas and Chemistry Analysis System Service Manual
Disassembling the Analyzer
9.4 Disassembling the Upper Assembly
9.4.1 Replacing the Touch Screen Drive Board
Follow the steps below to remove the touch screen drive board:
1. Separate the upper assembly and the lower assembly.
2. Disconnect all the wires on the touch screen drive board.
3. Remove the 4 screws securing the touch screen drive board.
4. Remove the touch screen drive board carefully.
To install a touch screen drive board:
Install a touch screen drive board in the reverse order. Refer to the following illustration when
- 77 -

85.

i15 Blood Gas and Chemistry Analysis System Service Manual
Disassembling the Analyzer
connecting wires, and ensure they are all correctly connected.
4
5
1
2
3
No.
Connect to
1
Touch screen
2
Lower assembly
3
LCD screen
4
Printer
5
Power button
9.4.2 Replacing the LCD
Follow the steps below to remove the LCD:
1. Separate the upper assembly and the lower assembly.
2. Remove the touch screen drive board.
3. Remove the 7 screws securing the LCD stander.
- 78 -

86.

i15 Blood Gas and Chemistry Analysis System Service Manual
Disassembling the Analyzer
LCD Stander
4. Remove the LCD stander, and place it on a flat surface.
5. Remove the LCD carefully.
To install a LCD:
Install a LCD in the reverse order. Ensure the wires are all correctly connected.
9.4.3 Replacing the Touch Screen
Follow the steps below to replace the touch screen:
- 79 -

87.

i15 Blood Gas and Chemistry Analysis System Service Manual
1.
Separate the upper assembly and the lower assembly.
2.
Remove the LCD.
3.
Remove the touch screen.
Disassembling the Analyzer
To install a touch screen:
Install a touch screen in the reverse order.
9.4.4 Replacing the Print Controller Board
Follow the steps below to remove the print controller board:
1. Separate the upper assembly and the lower assembly.
2. Disconnect all the wires on the print controller board.
3. Remove the screw securing the print controller board.
4. Remove the print controller board carefully.
To install a print controller board:
Install a print controller board in the reverse order. Refer to the following illustration when
connecting wires, and ensure they are all correctly connected.
- 80 -

88.

i15 Blood Gas and Chemistry Analysis System Service Manual
Disassembling the Analyzer
9.4.5 Replacing the Printer Head
Follow the steps below to remove the printer head:
1. Separate the upper assembly and the lower assembly.
2. Open the printer casing, and remove the 2 screws.
3. Pull the printer head carefully, and remove wires on the printer head.
To install a printer head:
Install a printer head in the reverse order. Refer to the following illustration when connecting
wires, and ensure they are all correctly connected.
- 81 -

89.

i15 Blood Gas and Chemistry Analysis System Service Manual
Replaceable Parts
10 Common Failures and Solutions
Error
Code
Prompts/Failure
Phenomenon
2
Cartridge Expired,
Please Change
Cartridge.
3
Cartridge Type Error.
4
Maybe Cartridge
Barcode Destroyed.
Scan Again or Try
Another Cartridge.
5
21
22
23
41
42
The Software Version
does Not Match with
Test Cartridge. Please
Update the Software
Immediately.
Calibrant Pack
Expired.Confirm Date
Setup and Calibrant
Pack's Expiration Date.
Causes
1) The test cartridge
expired.
2) The system date is
wrong.
The software doesn’t
support this kind of test
cartridge.
1) The barcode scanner
has been
contaminated.
2) The barcode has been
damaged.
The main control software
and the analog software
have not been updated.
1) The calibrant fluid
pack expired.
2) The system date is
wrong.
The software doesn’t
Calibrant Pack Type
support this kind of
Error, Please Replace it.
calibrant fluid pack.
1) The barcode scanner
Maybe Calibrant
has been
Barcode Destroyed,
contaminated.
Scan Again or Try
2) The barcode has been
Another.
damaged.
Quality Control
Solution Expired;
The controls expired.
Please Change Quality
Control Solution.
Quality Control
Solution Type Error.
Solutions
1) Change a new test
cartridge.
2) Correct the system date.
Update the analog software.
1) Clean the barcode
scanner.
2) Change a new test
cartridge.
Update the main control
software and the analog
software.
1) Change a new calibrant
fluid pack.
2) Correct system date.
Update analog software.
3) Clean the barcode
scanner.
4) Change a new calibrant
fluid pack.
Change the controls.
The controls test mode
doesn’t support the
Train the operator how to
selected kind of controls.
operate correctively.
Therefore, the system can’t
recognize its barcode.
- 82 -

90.

i15 Blood Gas and Chemistry Analysis System Service Manual
Replaceable Parts
Quality Control
Solution Level Error
The controls test mode
doesn’t support the
Train the operator how to
selected kind of controls.
operate correctively.
Therefore, the system can’t
recognize its barcode.
61
Current Date Error.
The calibrant fluid pack’s
valid period is inconsistent
with the system date.
Check whether the system
date is right. If not, correct
the date.
2005
Heating Abnormally,
Please Call for
Technical Assistance.
Heater failure
Change a new cam.
43
2008
Ambient Temperature
Out of Range
(10-31℃).
2009
Ambient pressure Out
of Range
(405-800mmHg).
2040
Front Board does Not
Contact Well with
Analog Board. Please
Call for Technical
Assistance.
1) The system’s
temperature is not
within the required
working temperature.
2) The system’s
temperature is within
the required working
temperature while the
temperature sensor
works improperly.
3) The system’s
temperature is within
the required working
temperature while the
wire connection is not
secure or correct.
1) The actual ambient
pressure is not within
the range.
2) The actual ambient
pressure is within the
range while the
pressure sensor has
some damages.
1) The connection wire is
not secured between
the front-end board
and the analog board.
2) The connection wire is
damaged between the
front-end board and
- 83 -
1) Ensure the system
works under the
required working
temperature.
2) Re-connect the wire.
3) Change a new wire.
1) Ensure the system
works within the
required pressure range.
2) Check whether the
pressure sensor has
been blocked.
3) Change a new analog
board.
1) Re-connect the wire.
2) Change the connection
wire between the
front-end board and the
analog board.

91.

i15 Blood Gas and Chemistry Analysis System Service Manual
Replaceable Parts
the analog board.
2050
Cartridge is Improperly
Ejected During Test
The test cartridge has not
been clicked into the
targeted place because of
improper operations.
2051
Calibrant Pack
Improper Removal,
Please Replace it.
Improper operation
2052
POGO_PIN does Not
Contact Well, Please
Try Another Cartridge.
1) Improper operation
2) The contact is poor
between the
POGOPIN and the
targeted part.
3) The electrical contacts
of the test cartridge
have been
contaminated.
2053
Chamber Door
Improper Open. Please
Close Chamber Door.
Improper operation
Close the chamber door.
Cartridge is Not
Qualified, Please Try
Another Cartridge.
1) The test cartridge has
been used.
2) There is some water
on the surface of the
electrical contacts
because the
1) Conduct test with a new
test cartridge.
2) Remove the water with
clean cloth from the
electrical contacts.
3) Train the operator not
2054
- 84 -
The operator needs to
perform the operation under
the requirements specified
in the User Manual.
Change calibrant fluid pack
under the requirements
specified in the User
Manual.
1) Re-conduct test with a
new test cartridge
following the
procedures described in
the User Manual.
2) Train the operator not
to contact the test
cartridge’s electrical
contacts to avoid
contamination.
3) If the external simulator
test succeeds, the
analyzer works
normally. If fails, the
internal simulator test
should be conducted. If
the internal simulator
test succeeds, check and
clean the POGOPIN. If
the internal simulator
test fails, change the
cam.

92.

i15 Blood Gas and Chemistry Analysis System Service Manual
3)
1)
2056
2064
Abnormal Cam
Location, Please Call
for Technical
Assistance.
Calibrant is Not
Aspirated Properly;
Please Try Another
Cartridge.
2)
3)
refrigerated test
cartridge has not been
returned to room
temperature as the
proper procedures or
the operator doesn’t
handle it properly.
The test cartridge’s
electrical contacts
have been
contaminated.
The motor can’t work
properly because the
poor connection of
wire.
Some structural parts
have been damaged.
The sensor can’t work
properly to detect the
corresponding
position.
1) The test cartridge has
some problems.
2) The calibrant fluid
pack needle’s plastic
washer is missing or
the pack has been
blocked.
3) The cam’s needle has
been blocked partially
or completely or
damaged.
4) The syringe pump has
been damaged which
will cause the failure
of calibration
aspiration absolutely.
- 85 -
Replaceable Parts
to directly contact the
electrical contacts.
1) Check whether the
structural parts have
been damaged or the
wire connection is not
secured.
2) Clean the sensor used to
detect the cam’s
position.
3) Reboot the analyzer for
testing.
1) If this failure occurs
occasionally, it’s
probably caused by test
cartridge. Change a new
test cartridge.
2) If this failure occurs
consistently, it’s
probably caused by
operator’s improper
operation. Remove the
calibrant fluid pack cap
and check whether the
plastic washer is
missing. Aspirate
calibration solutions
with syringe to
determine whether the
calibrant fluid pack is
blocking.
3) Check the needle which
is connected to the test
cartridge and located at

93.

i15 Blood Gas and Chemistry Analysis System Service Manual
Replaceable Parts
the cam assembly. If it
is contaminated, clean
it. If it is blocking or
damaged, please replace
it.
4) Change the syringe
pump.
2068
Sample Error, Maybe
Clots are Detected in
Sample.
1) The sample is not
1) The blood sample
connected into the test
should be correctly
cartridge.
connected into the test
2) Improper Operations
cartridge.
Don’t mix the
2) The operations should
blood sample
be correct.
timely after
The bubbles
collection; don’t
contained in the
discard the first
blood sample
drop of blood; the
should be removed
blood sample
and the blood
contains bubble.
sample after
The
collection should be
anti-coagulant
mixed timely.
heparin is not
The best blood
added into the
collection device is
blood sample.
the arterial blood
The blood sample
collection needle. If
contains
the arterial blood
anticoagulant
collection needle is
heparin but its
unavailable, use
concentration is
heparin as
not enough.
coagulant. We
The blood
recommend that two
sample’s volume
vials of 12500U/mL
is not enough. The
lithium heparin or
blood sample’s
heparin sodium
volume collected
should be added
by 1mL syringe is
into physiological
less than 0.5mL.
saline.
The blood
The volume of the
sample’s volume
collected blood
collected by 2mL
sample should be
syringe is less
enough.
- 86 -

94.

i15 Blood Gas and Chemistry Analysis System Service Manual
Replaceable Parts
than 0.8mL. The
blood sample’s
volume collected
by 5mL syringe is
less than 1.5mL.
2082
Insertion Error,
Cartridge is Inserted
during Simulator Test.
2084
Insertion Error,
Simulator is Inserted
during Sample Test.
2086
7001
Pump Sensor
Malfunctions, Please
Call for Technical
Assistance.
Input Current
Abnormal
7002
Battery Voltage
Abnormal
7003
5V Voltage Abnormal
7004
Printer Voltage
Abnormal
7005
12V Voltage Abnormal
Under electronic simulator
test mode, the test
cartridge can’t be inserted
into the analyzer.
1) Improper operation
2) The test cartridge is
contaminated leading
to a poor connection
1) Poor wire connection
causes that the motor
fails to work normally.
2) The motor’s screw
slips off.
3) There is a lot of dust
within the motor’s
pump.
4) The structural part is
damaged.
5) The sensor has some
problems and can’t
detect the
corresponding
position.
Re-conduct electronic
simulator test with a
simulator.
1) Operate properly
2) Change a new test
cartridge
1) Check whether the
structural part has been
damaged, whether the
wire connection is not
secured and whether the
screw slips off. If yes,
please fix it.
2) Remove the dust within
the motor’s pump.
3) Clean the pump and
check the sensor
The power adaptor or the
Change power adaptor or
battery has some problems. battery.
The battery has some
problems and the voltage
is abnormal.
If power voltage is
abnormal, the power
adaptor or the battery has
some problems.
The printer has some
problems and the voltage
is abnormal.
The power voltage is
abnormal, the power
- 87 -
Change battery or ensure the
analyzer is powered by
power adaptor.
Change the power adaptor
or the battery.
Change or fix printer.
Change the power adaptor
or the battery.

95.

i15 Blood Gas and Chemistry Analysis System Service Manual
7006
Battery Charge
Abnormal
USB Disk Cannot be
Identified
U disk Error
Can Not Get USB Disk
Information
Export Failure
Transmitting Failure
Backup Error
Printer Connection
Failure
IP Format Error
IP Conflict
adaptor or the battery has
some problems.
The battery has some
problems.
The USB Disk can’t be
recognized. The system
doesn’t support this type of
USB Disk.
The system can’t
recognize the USB disk.
The USB port has some
problems.
The function of USB disk
write protection has been
abled.
The internet cable is not
connected well; or the
network disconnects; or
the selected protocol is not
correct; or the server IP
address is not correct.
The targeted database is
empty or the USB disk is
turned off so that it can
only be read.
The model of the
connected printer is not
supported by the analyzer
or the USB port works
improperly.
The enter IP is 192.168.1.*
and * is not within 0-255.
The entered IP has been
used.
- 88 -
Replaceable Parts
Change the battery or reboot
the analyzer.
Use the correct type of the
USB Disk.
Format USB disk or change
a new USB disk.
Change the USB port or
read date via network port.
Disable the function of USB
disk write protection.
Connect the network cable
well and check the network
setup. Select the correct
protocol and set the correct
server IP address.
Disable the function of USB
disk write protection and
check that the database is
empty.
Only the printer with
designated model can be
connected to the analyzer.
Change the USB port or
reconnect USB cable.
Enter the correct IP. The
entered IP is 192.168.1.*
and * is within 0-255.
Re-enter a new IP. The
re-entered IP should not be
used or the IP can be
automatically obtained by
the analyzer.

96.

i15 Blood Gas and Chemistry Analysis System Service Manual
All Parameters Fail
Calibration; Please Try
Another Cartridge
1) The analyzer might
have some problems.
Please do external
simulator test.
2) The calibrant fluid
pack might have some
problems, such as
being stored not under
the required
conditions, being
contaminated or
introduced into blood.
3) The test cartridge’s
transport conditions
don’t meet
requirements.
4) The failure might be
caused by the
analyzer. The external
simulator test fails.
Please Reboot the
System, and Remove
Cartridge.
The barcode of test
cartridge has not been
scanned before the
cartridge has been inserted
into i15 analyzer.
Fail to Get Update File,
Please Check the
Network Setup
The server IP is not
192.168.1.93, or the tftpd
program has not been
opened, or the software
upgrading package is not
contained in the target file.
Update Failure
The network disconnects
when upgrading the analog
software.
- 89 -
Replaceable Parts
1) Repair cam assembly if
the external simulator
test fails.
2) Change calibrant fluid
pack.
3) Change test cartridge.
4) Store and use the test
cartridge and calibrant
fluid pack in
accordance with the
requirements listed in
the user manual.
1) Reboot i15 analyzer or
enter into Diagnostic
Screen to press Valve,
the cartridge will be
ejected out.
2) Require operator to
operate in accordance
with procedures
described in user
manual.
Check whether the network
setup is correct. Make sure
whether the server IP is
192.168.1.93. Make sure the
local IP is 192.168.1.* (*
can be any figure within
0-255). Open the tftpd
program. Place the software
upgrading package in the
target file.
Make sure the connected
network is stable when the
software is updated, or

97.

i15 Blood Gas and Chemistry Analysis System Service Manual
Replaceable Parts
update the analog software
via USB disk.
Replacement Failure.
Inter-Simulator Fails,
Perform External
Simulator Test
Power State is
Abnormal, Please Call
for Technical
Assistance
Barcode Check Error
The indicator on the
main screen doesn’t
The barcode of calibrant
1) Reboot the analyzer or
fluid pack can’t be scanned
scan the recognizable
successfully. The
barcode of the calibrant
installation of calibrant
fluid pack
fluid pack fails more than
2) Change the calibrant
two times.
fluid pack.
1) Conduct external
simulator test.
2) If the external simulator
test passes, disassemble
the cam assembly and
then check whether the
1) Cam’s wiring gets
zero-calibration board
loose.
is well connected, and
2) The POGOPIN of cam
finally re-assemble the
gets dampened.
cam.
3) The failure of front
3) If the external simulator
board
test fails, check whether
there are foreign objects
around the POGOPIN,
and dry up it and
re-conduct test.
4) Change Cam.
1) There are a lot of dust
on the power board,
1) Clean the analyzer and
which result in the
reboot it.
failure of the power
2) Re-install lithium
board.
battery.
2) The battery can’t work
3) Change lithium battery.
properly.
4) Change power board.
3) The power board can’t
detect the voltage.
1) The barcode is
1) Train the operator how
contaminated or
to operate properly and
damaged.
change the barcode.
2) The window of the
2) Clean the window of
scanner is
the scanner.
contaminated.
1) The connection of
1) Re-connect the power
power cord is poor.
cord.
- 90 -

98.

i15 Blood Gas and Chemistry Analysis System Service Manual
indicate the battery has
been charged,
The analyzer can’t be
powered by the battery
while can be powered
by AC power supply
normally.
The power adaptor
indicator is not on. The
analyzer can’t be
powered and the battery
can’t be charged.
The touch screen
doesn’t work
occasionally or
consistently.
Replaceable Parts
2) Failure of battery
2) Change the battery
Failure of battery
Change the battery
Failure of power adaptor
Change the power adaptor
1) Failure of touch screen 1) change touch screen
2) The screen board has
2) Change the screen
some problems
board
Fail to print
1) The installation of
printer paper is wrong.
2) The printer board
connection wire is
loose.
1) Correctly install printer
paper
2) Re-connect the printer
board connection wire
The DMS software
can’t be connected to
i15 analyzer.
Dongle’s information is
lost.
Reinstall DMS software
1) There are a lot of
unchecked sample
The data can’t be
information.
transmitted to the
2) The analyzer is
hospital’s system.
located in a place with
unstable and weak
WiFi signal.
DMS software can’t be The computer’s operating
opened.
system has some problems.
1) The database service
The system hints the
disconnects.
data has been
2) Other opened software
transmitted successfully
occupies 8000 port
while the DMS
such as other
software doesn’t
instrument’s central
receive the data.
station. Hospital
LIS/HIS, etc.
i15 analyzer can’t be
IP setting error
connected to the
- 91 -
1) Delete the unchecked
sample information.
2) Ensure the analyzer is
located in a place with
strong and stable WiFi
signal.
Re-install the computer’s
operating system.
1) Re-connect the database
service and re-transmit
the database.
2) Change the connection
port.
Reset the IP address

99.

i15 Blood Gas and Chemistry Analysis System Service Manual
computer.
The data can’t be
transmitted from i15
analyzer to DMS
software.
The data can’t be
transmitted to DMS
software.
DMS software hints the
test application time is
earlier than the test
inspection time.
The test cartridge’s
spring doesn’t work
normally.
The test cartridge has
solution leakage.
The calibrant fluid
pack’s chamber can’t
be opened.
The calibrant fluid pack
can’t be removed from
the analyzer.
Replaceable Parts
IP setting error
Reset the IP address
The DMS software needs
to be updated.
Update the DMS software
into the latest version
The test application time
should not be earlier than
the test inspection time
Reset the correct time
The test cartridge’s spring
is not in normal condition.
Reassemble the spring.
1) The test cartridge has
not been removed
immediately after
completion of test.
2) Improper operation
1) Check whether the
calibrant fluid pack’s
needle has blood, some
contamination or
blockade. If yes, change
the calibrant fluid pack.
2) Check whether the
POGOPIN has blood,
some contamination or
blockade. If yes, clean
it.
3) Change a new test
cartridge
The door of calibrant fluid
pack’s chamber has some
problems.
Change the door of calibrant
fluid pack’s chamber.
1) The key is not
removed from the
calibrant fluid pack
when installing the
pack.
2) Improper operation
1) Remove the test
cartridge firstly and
then turn off the
analyzer and finally
remove the calibrant
fluid pack from the
analyzer.
2) Instruct the operator to
correctly install the
calibrant fluid pack.
- 92 -

100.

i15 Blood Gas and Chemistry Analysis System Service Manual
The test cartridge can’t
be removed from the
analyzer occasionally.
The test cartridge can’t
be inserted into the
targeted position from
the analyzer’s port and
it will be ejected out
automatically.
The system halts when
its data exported to the
USB disk.
i15 analyzer can’t be
switched off.
The spring which is
located in the cam and
used to pull out the test
cartridge after completion
of test has some problems.
main control software’s
failure
main control software’s
failure
2)
1)
QC tests with controls
fail.
Change a new spring.
Improper operation and the Normalize the operation and
test cartridge doesn’t click ensure the test cartridge
into place.
clicks into place.
1)
The external simulator
test fails.
Replaceable Parts
2)
3)
4)
Upgrade main control
software
Upgrade main control
software
1) Conduct internal
The POGO PIN has
simulator test
some foreign matter
2) If the internal simulator
which causes the
test succeeds, check
failure of external
whether the POGO PIN
simulator test
has some foreign matter
The channel
and clean the
inspection results
POGOPIN
don’t meet
3) If the internal simulator
requirements.
test fails, change the
cam.
The scanned barcode
1) Scan the barcode of
is not the controls’
controls used for the
barcode used for the
test.
test.
2) Conduct tests with the
Don’t conduct tests
proper controls
with the proper
3) Re-conduct test with a
controls.
new calibrant fluid
The calibrant solutions
pack.
are contaminated.
4) Conduct simulator test.
The simulator test may
If the test fails, maintain
be out of control.
or replace the cam.
- 93 -

101.

i15 Blood Gas and Chemistry Analysis System Service Manual
The test cartridge can’t
be recognized after its
barcode has been
scanned.
1) The test cartridge
doesn’t click into
place because of the
operator’s improper
operation.
2) The test cartridge has
some problems.
It’s difficult to scan the
test cartridge’s barcode.
1) The operator doesn’t
do the operation in
accordance with
standard procedures.
2) The barcode is not
clean enough.
The system hints to
remove the test
cartridge after being
logged while there is no
test cartridge actually.
1) The detection slot of
the test cartridge has
some foreign objects.
Therefore the test
cartridge sensor can’t
work normally.
2) The wiring connection
is poor.
- 94 -
Replaceable Parts
1) The operator needs to
operate in accordance
with the user manual.
2) Check whether the
detection slot of the test
cartridge has some
foreign objects. If yes,
remove the foreign
objects.
3) Check whether wires
connected to the
detection board of the
test cartridge become
loose. Reconnect the
FPC board which
detects pump.
1) Train the operator how
to operate correctly.
2) The distance between
the barcode and scanner
should be within
10-15cm.
3) Contact manufacture
1) Check whether the
detection slot of the test
cartridge has some
foreign objects. If yes,
remove the foreign
objects.
2) Check whether wires
connected to the
detection board of the
test cartridge become
loose. Reconnect the
FPC board which
detects pump.
3) Replace FPC board
which detects pump.

102.

i15 Blood Gas and Chemistry Analysis System Service Manual
HCT AC impedance is
too high.
1) The electrical contacts
of test cartridge are
contaminated.
2) The electronic
simulator test may be
out of control and the
POGOPIN may be
contaminated.
3) There are bubbles
when calibrant
solutions are aspirated
into the test cartridge.
This case may be
caused by the
following reasons:
Occasional leaks
or blocking of test
cartridge
The calibrant fluid
pack is blocked
partially.
The valve
connecting
analyzer and
calibrant fluid
pack or test
cartridge is not
positioned in a
correction
position.
4) The occasional failure
of test cartridge’s Hct
sensor.
The calibration
succeeds while one or
several items have no
test results.
The volume of blood
sample is not enough or
there are bubbles in the
blood sample.
pH has no test result
occasionally or other
parameters have no test
results occasionally.
1) The occasional failure
of test cartridge.
2) The sample has some
problems.
- 95 -
Replaceable Parts
1) Train the customers not
to touch the electrical
contacts of test
cartridge to avoid
contamination.
2) Conduct electronic
simulator test. If the test
fails, firstly clean the
POGOPIN, then
re-conduct the
electronic simulator
test. If the test fails
again, the cam needs to
be replaced.
3) Re-conduct tests with a
new test cartridge;
Re-conduct tests with a
new calibrant fluid
pack; Re-conduct tests
after the valve has been
positioned correctly.
4) Re-conduct test with a
new test cartridge
Train the personnel who is
responsible for sample
collection to normalize the
procedures of sample
collection and test.
1) Re-conduct test with a
new test cartridge.
2) Train the personnel who
is responsible for
sample collection to

103.

i15 Blood Gas and Chemistry Analysis System Service Manual
The test results are
displayed with ***
Controls test fails.
The barcode scanner
doesn’t work.
The scanner’s wire
connection is loose.
- 96 -
Replaceable Parts
normalize the
procedures of sample
collection and test.
1) Re-conduct QC test.
2) Train the personnel who
is responsible for
sample collection.
Re-connect the scanner’s
wire,

104.

i15 Blood Gas and Chemistry Analysis System Service Manual
Replaceable Parts
11 Replaceable Parts
WARNING
Only the replaceable parts supplied by EDAN or its authorized distributors should be
connected to the analyzer.
Part
Part Number
Main Control Board PCBA
02.03.451152-11
Power Board PCBA
02.02.451154-11
Analog Board PCBA
02.03.451128-11
Touch Screen Drive Board PCBA
02.02.451532-11
Wi-Fi Board PCBA
12.02.451363
Chamber Door Detection Board PCBA
02.02.451663-11
Power Interface Board PCBA
02.02.451357-11
Ambient Temperature Detection Board
02.02.451751-11
Cam Machine
02.01.210726
Speaker
01.14.038010
Exhaust Fan
21.58.472029
LCD
01.16.045068
Touch Screen
21.16.045069
Bar Code Scanner
21.17.047380
Print Controller Board
02.02.220047-11
Printer Head
01.17.078173
- 97 -

105.

- 98 -
English     Русский Rules