1. Overview_V2.0

1.

Overview
H-120 MCSR Training

2.

Course Overview
Course objectives:
Through the study and assessment of this course, students can independently complete
the active and passive service of H 120 products, including installation, troubleshooting,
maintenance and calibration.
Course composition:
H 120 Overview
H 120 Learn about the product (clinical knowledge & measurement principles; Fluidic
System; Hardware System; Mechanical System)
3
Adjustment
Maintenance
Installation
Calibration and QC
Troubleshooting
©2022 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All
rights Reserved.

3.

H 120/H120P Instrument overview

4.

10 years of precipitation, technological breakthrough, strong attack!
©2024 Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
All rights Reserved.
H-120 Automated Glycohemoglobin Analyzer

5.

Instrument Model and Measurement Parameters
©2024 Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
All rights Reserved.
Product Description: This product is based on the principle of high performance liquid chromatography
HPLC . It is used together with the supporting reagent and chromatography column to quantitatively
detect HbA1c in human blood samples.
Model
Storage Sample Volume
Description
H-120
100000 test results
H-120P
50000 test results
Except for the storage space difference, the
operating modes, output parameters and
reagent/consumables of the two models are the
same.
Name
Parameter
Abbreviat
ion
Unit
Hemoglobin A1c
HbA1c
HbA1c
%; Mmol/mol
Fetal hemoglobin *
HbF
HbF
%
Total glycosylated
hemoglobin *
HbA1
HbA1
%
Average Blood Sugar *
eAG
eAG
Mmol/L Mg/dL
* research use only parameters.

6.

Measurement mode
©2024 Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
All rights Reserved.
• Test Mode
test panel
Standard mode
Variant mode
• Blood mode
Speed
(samples/hour)
120
60
Analyzer alarm
capability
Suspected variant E
Suspected variant U
Suspected variant BK
Suspected variant
Suspected variant S
Suspected variant C
Suspected variant D
Suspected variant E
Suspected variant U
Suspected variant BK
Other suspected
variants
Feeding position
Auto
Closed
Blood mode
WB
WB
PD (by DI water)
• Sample Volume
WB/Micro-WB 16ul
PD 10ul

7.

Reagent
©2024 Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
All rights Reserved.
Chinese name
Model and
Storage
Specification
condition
Storage
validity
period
(Months)
Open-vial
validity
period
(Days)
Eluent A for HbA1c analysis
H-12 A:1500 ml
2-30 ℃
24
100
Eluent B for HbA1c analysis
H-12B:500 ml. 250 ml
2-30 ℃
24
100
Hemolysis Solution
H-12 H:5600 ml
2-30 ℃
24
100
2-8 ℃
12
14
12
7
HbA1c control (same as H50)
2mL*2 vials/box
2mL*1 vials/box
HbA1c calibrator (same as H50)
2mL*2 vials/box
2-8 ℃
HbA1c chromatographic column
H-3 : 3000 test
H-6 : 6000 test
H-10 : 10000 test
2-8 ℃
Filter
3000 test
2-8 ℃
24
/
Probe cleanser
50ml
2-30 ℃
12
60
12
Eluent A
Eluent B
/
Relationship between the chromatography column model and filter accessories
Model
H-3 3000 times
Accessories
Number of filters: 1
H-6 6000 times
H-10 10000 times
Quantity of filters: 2
Number of filters: 4
Lyse

8.

Easy operation and easy replacement of reagent
©2024 Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
All rights Reserved.
column&filter
Replace the column&filter
RFID technology, automatic identification
Open, take, set, close, 5s, four steps, senseless replacement,
Time and effort saving
Indicator, color change prompt status
Eluent A & Eluent B
Replace Eluent A&B
• RFID technology, automatic identification
Reagent status
• Quick plug-and-pull, one-step in-place
• Indicator, color change prompt status

9.

Basic operation
©2024 Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
All rights Reserved.
Touch screen, It displays all alphanumeric
and graphic data you may need to operate
your analyzer.
[Mode Switch] key: Press the
[Mode Switch] key to switch sample
presentation mode.
Count key: Press the [Count]
key to start count or Probe
Cleanser maintenance.
Autoloader 10 racks X 10 tube positions
5 racks X 10 tube positions
STAT position, supports quick sample loading
Closed-tube sampling mode.
The analyzer supports all types of sample
handheld tests
• Steady blue: ready
Status
indicator
• Flashing blue: running
• Steady red: in error
• Steady orange: standby
• Off: shutdown

10.

Instrument Function Introduction
1. Sample options :
used for sample test information editing,
measurement mode selection and result display
2. Reagent options :
displays reagent expiration date and other usage
information
3. Quality control options :
automatic quality control analysis function
4. Calibration options :
automatic calibration function
5. Maintenance options :
instrument probe fluid maintenance, whole machine
cleaning, filling and other functions
6. Service options :
view instrument status, version information and
other functions
7. Settings options :
schedule instrument power on/off time, instrument
parameters and other settings
©2024 Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
All rights Reserved.
1
2
3
4
5
6
7
Main interface of the instrument

11.

Analyze Screen-Sample Program
1. click Sample
Order
2. Select
Sample type
3. Select mode
4. Sample
numbering
5. Select OK
©2024 Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
All rights Reserved.

12.

Analysis Screen – Result Display
serial
number
©2024 Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
All rights Reserved.
details
Report parameters
1.Glycated hemoglobin HbA1c
1
Research parameters
1.Fetal haemoglobin HbF
2.Total glycosylated haemoglobin
HbA1a%+HbA1b%+HbA1c%
3.Estimated average glucose eAG
Chromatographic peak results
display
2
left to right Peak-retention timePeakArea-Percentage of areaReportable results
1
2

13.

Analysis Screen – Chromatogram
serial
number
©2024 Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
All rights Reserved.
details
chromatogram left to right
3
Peak1 A1a
Peak2 A1b
Peak3 Hb F
Peak4 L-A1c
Peak5 S-A1c
Peak6 HbA0
Peak7 Variant
Peak8 System peak
Flag
4
1.abnormal results flags
2.abnormal chromatogram flags
3.variant flags
Peak time in standard mode:
HbA1cRT=13 s±0.5
A0 RT=21s±1s
Peak time in variant mode:
HbA1cRT=17s±0.5
A0 RT=31s±1s
4
SA1c
System
peak
L-A1c
A1b
A1a HbF
Hb A0
3
Variant

14.

Power-on/off function
©2024 Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
All rights Reserved.
When the main power of the analyzer is switched off, turn on the main power switch at the back of the
analyzer.
Click the shutdown button on the upper right corner of the analyzer software screen to perform the
shutdown procedure. After the procedure is completed, the power will be switched off automatically, and
there is no need to manually turn off the main power of the analyzer.
When the analyzer is shut down according to the shutdown procedure and the main power of the
analyzer is on, you can press the key on the right side of the analyzer screen to turn on the analyzer
(recommended).
Key on the right side of the screen
Main power supply

15.

Auto Startup and Shutdown
• The analyzer provides the auto startup/shutdown function, and is set on the
analyzer screen.
• To use the auto startup function, do not power off the analyzer main power
switch.
©2024 Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
All rights Reserved.

16.

Summary
Remember the basic performance of H-120: Test mode, test speed, test parameters, and
reagent
17
Learn how to replace reagent (highlight)
Learn the basic operation of the instrument and composition of the main screen
Learn the soft/hard shutdown mode and auto startup/shutdown mode
©2022 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All
rights Reserved.

17.

Thank you!